![M-system 47LV Скачать руководство пользователя страница 18](http://html2.mh-extra.com/html/m-system/47lv/47lv_operating-manual_3101030018.webp)
47LV
P. 18 / 18
EM-9502-B Rev.3
M-SYSTEM WARRANTY
M-System warrants such new M-System product which it manufactures to be free from defects in materials and workmanship during the 36-month period following the date that such
product was originally purchased if such product has been used under normal operating conditions and properly maintained, M-System’s sole liability, and purchaser’s exclusive
remedies, under this warranty are, at M-System’s option, the repair, replacement or refund of the purchase price of any M-System product which is defective under the terms of this
warranty. To submit a claim under this warranty, the purchaser must return, at its expense, the defective M-System product to the below address together with a copy of its original
sales invoice.
THIS IS THE ONLY WARRANTY APPLICABLE TO M-SYSTEM PRODUCT AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. M-SYSTEM SHALL HAVE NO LIABILITY FOR CONSEQUENTIAL, INCIDENTAL OR
SPECIAL DAMAGES OF ANY KIND WHATSOEVER.
M-System Co., Ltd., 5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN, Phone: (06) 6659-8201, Fax: (06) 6659-8510, E-mail: [email protected]
ERROR MESSAGES
MAIN DISPLAY
Input error, Out of the measuring range
Non-volatile memory error (reading)
*
1. If the unit does not recover its function after the initialization, repairing in the factory may be required.
Non-volatile memory error (writing)
Increase/decrease the input signal until it is back within the measuring range.
While the error message is on the display, press Up button for 3 seconds or
more, go to the lockout setting mode and initialize the unit to its factory
default status.
*
1
Repair is needed if the display does not recover after the power is reset.
ERROR MESSAGE
WHAT TO DO
Internal data error
CHARACTER SET
0 1 2
4
3
5 6 7
9
8
– A B
D
C
E F G
I
H
J K L
N
M
O P Q
S
R
T U V W X
Z
Y
■
LOOP TEST OUTPUT MODE
2.00
Shift
Loop Test Output
Measuring Mode
-19999...19999
Set value
Setting Parameters
Hold down Alarm/ + Scale/ + Shift
button for ≥5 seconds
Hold down Alarm/ or Scale/ button
for ≥1 second
Loop Test Output Mode
*
1. Hold down Alarm/ or Scale/ button for ≥1 second to return to the measuring mode while setting parameters.
*
1
PARAMETER
Loopt test output: Pressing [Shift] button switches the signal to increase ( ON) or decrease ( ON), and then pressing [Up]
button controls it toward the desired output value.
Alarm trip functions according to the scaling values during the loop test.
*
1. The specified decimal point position is applied to the loop test output value. -19999 to 19999 when ‘No decimal fraction’ is specified.
INDICATORS
/
FUNCTION
DEFAULT
VALUE
DISPLAY
Tch
Fnc
Min
Max
...
Loop test output
Scaling value for the loop test output
(display blinking)
Zro
Spn D/P
N/A
Max
Min
*
1
LIGHTNING SURGE PROTECTION
M-System offers a series of lightning surge protectors for
protection against induced lightning surges. Please contact
M-System to choose appropriate models.