47LV

P. 10 / 18

EM-9502-B Rev.3

PARAMETER

INDICATORS

FUNCTION

DEFAULT

VALUE

DISPLAY

*

1. 

During setting with parameters, ‘Tch’ turns off.

*

2.

 

‘Default Value’ shows values for input code in order from 1 through 6.

Analog output functioin mode: Forced zero, low-end cutout and display refreshing rate are applied to the display value, thus affecting 
 

the analog output when the mode is set to ‘Proportional to the display value.’

 

The input signal directly affects the analog output with ‘Proportional to the scaling value’ setting.

Analog output 0% / 100% adjustment: Pressing [Shift] button switches the signal to increase or decrease, and then pressing [Up] 
 button controls it toward the desired output value.  Analog output 0% + 5% ≤ Analog output 100%

Analog output 
function mode

Proportional to the display value

Fnc

Zro

Spn Fnc

Proportional to the scaling value

Analog output 0%
adjustment

Analog output 0% adjustment:
adjustable range -5 to 100%

(increasing)
(decreasing)
(increasing)
(decreasing)

Fnc

Analog output 100%
adjustment

2.00

2.00

2.00

Analog output 100% adjustment:
adjustable range 0 to 105%

0% output

(default value)

100% output

(default value)

A

A

B

B

Display

Zero

Full Scale

Input

Input Scaling

Display Scaling

B

A

B

A

Display

Full Scale

Zero

Input

Input Scaling

Display Scaling

• Normal Scaling

The display value increases when the input signal increases.

• In

ve

rted Scaling

The display value decreases when the input signal increases.

Decimal point position

*

2

4 positions or none

Decimal point position (10

–4

 

2.00

)

D/P

‘Teach’ Calibration

‘Input scaling value A’ and ‘Input scaling value B’ can be adjusted by applying actual input signals.  These settings will be 
lost after an initialization.

How to ‘Teach’ Input Values

1)  Hold down Scale/

 button for 3 seconds or more to enter the scaling setting mode.

2)  Choose an appropriate input type and press Alarm/  button to move on to Input scaling value A.
3)  ‘Zro’ and ‘Tch’ indicators turn on.
4)  Zero point: Press Up button and ‘Tch’ indicator starts blinking.
5)  Apply 0% input signal.  Press Up button again and ‘Tch’ indicators returns to ON.
6)  Press Alarm/  button to move on to Display scaling value A.
7)  Press Alarm/

 button again to move on to Input scaling value B.

8)  ‘Spn’ and ‘Tch’ indicators turn on.
9)  Full-scale point: Press Up button and ‘Tch’ indicator starts blinking.

10

) Apply 100% input signal.  Press Up button again and ‘Tch’ indicators returns to ON.

Содержание 47LV

Страница 1: ...y indicate it POWER INPUT RATING OPERATIONAL RANGE Locate the power input rating marked on the product and confirm its operational range as indicated below 100 240V AC rating 85 264V 50 60 Hz approx 6...

Страница 2: ...ripped HH turns on when the HH alarm is tripped P turns on when none of the other alarms is tripped 3 Function indicators Indicate the status in each setting mode 4 Max Min button Used to switch the m...

Страница 3: ...THE UNIT ON A PANEL The watertight packing must be in place to hold the meter Do not remove it 1 Insert the unit into the panel cutout 2 Push the mounting brackets into the grooves on both sides of t...

Страница 4: ...Lb NC La NO Hc COM Hb NC Ha NO V 17 15 16 ALARM SUFFIX CODE 2 SPDT contact 2 points 20 18 19 Lc LLc LLa La Hc HHc Ha HHa 2 3 4 INPUT LL OUTPUT L OUTPUT H OUTPUT L OUTPUT H OUTPUT HH OUTPUT 1 9 10 POWE...

Страница 5: ...Hold down Max Min Alarm buttons at once for a preset time duration 1 Hold down Alarm Scale buttons at once for 3 seconds POWER ON 1 INITIAL SETTING FLOWCHART Initial Setting Complete Measuring Starte...

Страница 6: ...Zero cannot be executed when the operation is prohibited with the lockout protection setting MAX MIN display mode cannot be cancelled while in the Forced Zero mode Forced Zero cannot be cancelled whi...

Страница 7: ...ess Up button to change the blinking value Press Shift button to go to the next digit Press Alarm or Scale button to apply the new value and go to the next or previous parameter setting Invalid parame...

Страница 8: ...on for 1 second Shift Shift Up Up Up set Up set Alarm Scale Alarm Alarm Scale Alarm Alarm Scale Alarm Scale Alarm Shift Shift Alarm Scale Alarm Scaling Setting Mode Alarm Alarm Alarm 1 Alarm Scale Sca...

Страница 9: ...to 100 A Measuring range 200 to 200V Measuring range 1 to 1mA Measuring range 10 to 10mV Measuring range 100 to 100mV Measuring range 1 to 1V INDICATORS FUNCTION DEFAULT VALUE DISPLAY Zro Spn Zro Tch...

Страница 10: ...default value 100 output default value A A B B Display Zero Full Scale Input Input Scaling Display Scaling B A B A Display Full Scale Zero Input Input Scaling Display Scaling Normal Scaling The displ...

Страница 11: ...larm Setting Mode Alarm Shift L Setpoint L Deadband L ON Delay Time 19999 19999 0000 9999 00 99 Shift Shift Alarm Alarm Alarm Alarm Alarm Alarm L Coil at Alarm RYEN RYDN Shift Alarm Alarm Shift L Trip...

Страница 12: ...B HH ON Delay Time 00 99 Shift Alarm Alarm Alarm Alarm HH Coil at Alarm RYEN RYDN Shift Alarm Alarm Shift HH Setpoint 19999 19999 Shift H Setpoint H Deadband H ON Delay Time 19999 19999 0000 9999 00...

Страница 13: ...Hi trip L Lo trip L coil at alarm L Coil energized at alarm L Coil de energized at alarm H setpoint H deadband hysteresis H ON delay time H ON delay time seconds H Deadband hysteresis value H Setpoint...

Страница 14: ...the scaled range values Alarm is disabled when is specified for the setpoint Note 4 All alarm setpoints are disabled reset to status when the input type has been changed Re setting is required Note 5...

Страница 15: ...tting Parameters Shift Scale Alarm Low End Cutout Low End Cutout Value OFF ON 00 99 Shift Alarm Scale Alarm Alarm Scale Scale Scale Scale Scale Scale Scale Scale Scale Scale Scale Scale Scale Alarm ou...

Страница 16: ...m the setting mode Transition time to Lockout Setting Mode The display goes to Lockout Setting Mode only when the designated buttons are pressed for the specified time duration 1 F00 0 25 msec refresh...

Страница 17: ...Setting Mode in Advanced Setting Mode A Scale Scale Scale Scale Scale Scale Set value Setting Parameters A Forced Zero Lockout OFF ON Shift Alarm Alarm PARAMETER Alarm setting lockout Lock Alarm Sett...

Страница 18: ...repairing in the factory may be required Non volatile memory error writing Increase decrease the input signal until it is back within the measuring range While the error message is on the display pres...

Отзывы: