Referencing Standard Depth Gauge 

Referencing Depth Gauge with hook

1. Ensure the locking screw is loose.

2. Wipe clean surface plate or similar flat 

service.

3. Ensure Gauge base is clean.

4. Place Gauge base on the surface plate.

5. Slide the beam downwards so 

measuring contact is in contact with the 

surface plate.

6. Press the zero button.

1. Ensure the locking screw is loose.

2. Ensure the Gauge base and hook is 

clean.

3. Close the Gauge.

4. Press the zero button.

Metric/Inch Conversion

25.40

mm

in/mm

Press in/mm to convert from 

inch to metric display.

1.000

The display will indicate 

which units are being 

displayed.

in/mm

in

Trouble Shooting

Problem

Possible Cause

Solution

Flashing display

Low voltage

Poor contact 

Replace battery

Clean contact

No display

Dead or weak battery

Poor contact 

Replace battery

Clean contact

Erratic or frozen display

Contamination between beam & head

low voltage

Wipe Depth Gauge clean

Replace battery

Erroneous reading

Low voltage

Contamination between beam & head

Too much operating force

Low ambient temperature

Replace battery 

Wipe Depth Gauge clean

Use pressure device

Store Depth Gauge at normal room 

temperature