setting mode 

– signalled with (d.br) notice. 

 
In this mode, by means of “+” and “-“ buttons the user has the possibility to set the brightness of the display 
segments.  Exit  from  the  mode  and  confirming  the  setting  occurs  when  the  “ON/OFF”  button  is  pressed 
shortly.  Longer  pressing  and  holding  of  the  button  and  subsequent  releasing  of  the  button  will  activate  the 
calibration mode, signalled by the (CAL) notice. In this mode by means of “+” and “-“ buttons, the user has a 
possibility  to  adjust  the  temperature  readings  to  the  real  temperature.  Exist  from  the  mode  and  confirming 
the calibration settings occurs when the “ON/OFF” button is pressed shortly.  
 

NOTE 

– the controllers supplied have been calibrated already. 

 

 

Longer pressing and holding of the button and subsequent  its releasing shall activate the preliminary heating 
time  setting  mode,  signalled  by  (P.tl)  notice.  In  this  mode  by  means  of  “+”  and  “-“  buttons  the  user  has  a 
possibility  to  define    the  time  after  which  the  controller  gets  switched  from  preliminary  heating  phase  to 
proper  heating  phase.  Deactivation  of  preliminary  heating  is  signalled  by  the  “OFF”  notice.  Exist  from  the 
mode and setting confirmation occurs when the “ON/OFF” button is pressed shortly.  
 
Longer pressing and holding of the button and its subsequent releasing will activate the preliminary heating 
temperature setting mode, signalled by  (P.tE.) notice. In this mode by means of “+” and “-“ buttons the user 
has  a  possibility  to  define  the  preset  temperature  that  shall  be  maintained  during  preliminary  heating.  Exit 
from  the  mode  and  setting  conformation  occurs  when  the  “ON/OFF”  button  is  pressed  shortly.      Longer 
pressing  and  holding  of  the  button  and  its  subsequent  releasing    will  activate  the  preset  temperature  limit 
setting  mode,  signalled  by    (L.t.h.) 

notice.  In  this  mode  by  means  of  “+”  and  “-“  buttons  the  user  has  the 

possibility  to  set  the  upper  limit  of  the  preset  temperature  settings.  Exit  from  the  mode  and  setting 
confirmation occurs after the “ON/OFF” button is pressed shortly 

 

NOTE – all controller settings and working state (activated or deactivated) are stored in the 
non-volatile memory. 
 

Entering the access codes 

 

During controller start-

up (displayed controller’s name, software version, settings), press and hold the “+” and 

“-“ buttons.  Once “---“ has been displayed on the screen, buttons may be released and the relevant code 
can be set. The code shall be confirmed by the “ON/OFF” button. 

 

CODE 

ACCESS LEVEL 

Any 

L-0 

157 

L-1 

314 

L-2 

628 

L-3 

942 

L-4 

 

Working time setting mode (code L-0)

 

Display brightness setting mode (code L-0)

 

Calibration mode (code L-1)

 

Preliminary heating time setting mode (code L-2)

 

Preliminary heating temperature setting mode (code L-3)

 

Preset temperature limit setting mode (code L-4)

 

 

     Controller error report 

 

HC1 controller has been equipped with advanced mechanisms for error detection. Detection of any 

error  activates  emergency  work stoppage  and  triggers  error  report  screen. Error  report  screen  is  displayed  
in a continuous manner. It is therefore necessary to  disconnect power supply, remove the error source and 
controller reconnection 

 

ERROR 

ERROR DESCRIPTION 

(E-0) CPU 

STATUS 

Damaging the main processing unit. 

Содержание W204100A

Страница 1: ...H x D Temperature sensor HC 01 HC 01 Number of shelves 3 2 Fan 230V 14 Wat 230V 14 Wat Heating element 80Wat 80Wat The adjustable fan anemostat is fixed on the top of the chamber A thermal switch off...

Страница 2: ...ange 30 95 C Temperature range 30 95 C 30 95 C Internal dimensions 690x1010x740 mm W x H x D Internal dimensions 910x1210x1300 mm W x H x D 1310x1160x130 0 mm W x H x D External dimensions 875x1205x86...

Страница 3: ...lator switched on and initial heating in progress 2 signalling the activation of heating transmitter Indicator lights up transmitter contacts closed heating on indicator dimmed contacts opened heating...

Страница 4: ...phase to proper heating phase Preliminary heating temperature setting mode pressing the button will decrease the value of preset temperature that will be maintained during preliminary heating Preset t...

Страница 5: ...shall be maintained during preliminary heating Exit from the mode and setting conformation occurs when the ON OFF button is pressed shortly Longer pressing and holding of the button and its subsequent...

Страница 6: ...lue of preset temperature 30 C Maximum value of preset temperature Set up in the range 45oC 95oC Setting range for automatic deactivation 1 96 hours Setting range for preliminary heating temperature 3...

Страница 7: ...dedicated feeder Measurement input for temperature measurement PT1000 Outlet type Relay contact NO Output load AC1 9A 230V Brush strips sealing the bottom chamber are numbered and attached to the cham...

Страница 8: ...x g boko 550x810x440 mm szer x wys x g boko Czujnik temperatury HC 01 HC 01 Ilo p ek 3 2 Wentylator 230V 14 Wat 230V 14 Wat Grza ka 80Wat 80Wat Regulowany wentylator anemostat montowany jest od g ry...

Страница 9: ...Regulacja temperatury 30 95 C 30 95 C Wymiar wewn trzny 690x1010x740 mm szer x wys x g boko Wymiar wewn trzny 910x1210x1300 mm szer x wys x g boko 1310x1160x13 00 mm szer x wys x g boko Wymiar zewn t...

Страница 10: ...pnego 2 sygnalizacja w czenia przeka nika grzania Wska nik wieci styki przeka nika zwarte grzanie w czone wska nik nie wieci styki rozwarte grzanie wy czone 3 wy wietlacz Tryb pracy domy lny tryb samo...

Страница 11: ...y warto maksymalnej temperatury zadanej jak b dzie mo na nastawi 5 przycisk zwi kszenia warto ci Tryb pracy Naci ni cie przycisku zwi kszy warto temperatury zadanej Podczas nagrzewania wst pnego mo li...

Страница 12: ...ywaj c przycisk w oraz u ytkownik ma mo liwo ustawienia g rnego ograniczenia nastawy temperatury zadanej Wyj cie z tego trybu oraz zatwierdzenie nastawy nast puje po kr tkotrwa ym naci ni ciu przycisk...

Страница 13: ...ulacji Dwustanowa Parametry elektryczne sterownika Zasilanie p ytki sterownika Zasilanie dedykowanego zasilacza 12VDC 10 Min 200mA 100 240VAC 50 60Hz Wej cie pomiaru temperatury PT1000 Typ wyj cia Prz...

Страница 14: ...A B A B A A otwory monta owe listew B oznaczenie listwy i boku np 4...

Страница 15: ...65 90 65 16 900 90 18 900 100 100 30 55 C 30 55 C 340x520x320 x x 380x540x340 x x 500x740x520 x x 550x810x440 x x HC 01 HC 01 3 2 230 14 230 14 80 80 80 4 4 490 4...

Страница 16: ...W40780 W40781 490 300 120x120 2 2 2 30 55 C 30 55 C 30 55 C 690x1010x740 x x 910x1210x1300 x x 1310x1160x13 00 x x 875x1205x865 x x 1135x1290x1535 x x 1520x1235x15 35 x x HC 01 HC 01 HC 01 4 0 0 230...

Страница 17: ...01 Heating Controller c 4 5 6 3 1 2 3 1 1 2 0 1 0 3 0 1 0 1 4 5 Prog 3 2 1 2 3 o C o C Pro ON OFF d br ON OFF...

Страница 18: ...CAL L 1 ON OFF o C P tI L 2 ON OFF OFF HC2 P tE L 3 ON OFF P o C L t h L 4 ON OFF L o C 4 5 6 ON OFF ON OFF OFF ON Pro OFF...

Страница 19: ...ON OFF d br ON OFF CAL ON OFF P tI OFF ON OFF P tE ON OFF L t h ON OFF ON OFF L 0 157 L 1 314 L 2 628 L 3 942 L 4 L 0 L 0 L 1 L 2 L 3 L 4 3 3 HC1...

Страница 20: ...0 CPU STATUS E 3 T Tmin T1 E 4 T Tmax T1 E 5 Przycisk E 6 Przycisk E 7 Przycisk ON OFF ON OFF 3 4 FW 0 1 50o C 250o C 0 1 C 1 5 C 30 C 45o C 95o C 1 96 30o C 40o C 0 60 12VDC 10 200 A 100 240VAC 50 6...

Страница 21: ...PT1000 NO AC1 9A 230 2000 230VAC 3 x 104 10A 230VAC AC1 600 0o C 55o C 0o C 60o C 65 25 o C A B A B A...

Страница 22: ...A B 4...

Отзывы: