Working Mode 

 

1. 

Press the power switches of the two scales to turn on the power of the scales.

 

 

 
 
 

2.

 

Press [ON/OFF] key on the indicator to turn on the indicator. The bluetooth LEDS on 

the indicator and the scale body both change from flashing to bright, indicating that the 

indicator and the scale body are connected successfully. The indicator will display a 

string of self-

checking characters from ‘PL===’ to  “PL---” “Pr---”“33333” for a few 

seconds and “------”then 0.00, now the scale enter the WEIGH mode. 

 

   

Cautions 

 

1.  if you press the [on/off] key on the indicator to turn on the indicator, the bluetooth 

indicator on the indicator will always flash, and the indicator will always display the 

self-test character of "PL = = ="or  "Pr = = ="

 

  

The reason for this phenomenon may be that the power switch of  the  left scale 

body or right scale body 

isn’t turned on. Another possibility is that the battery of the 

scale body is too low, please charge in time. 

2

When the sign “

” is lit, it indicates the indicator is low battery.  

When the sign “

” is Slow flicker it indicates the left scale body is low battery. 

 

When the sign “

” is

 Fast flashing, it indicates the right scale body is low battery.

 

Please charge the battery for 4 to 6 hours. When the charging LED turns from 

yellow to green, it indicates that the battery charging is completed. Please unplug 

the charger (adapter) in time, because overcharging will cause battery damage.  

3. Only use the charger (adapter) supplied with the scale and indicator. 

4.After charging, the battery does not last long, please contact the supplier to replace 

new battery. 

 
 

 

Содержание Tree LWC-M 400

Страница 1: ...rofessional Weighing Equipment Wireless Wheelchair Scale Operation Manual LWC P Series NOTE THIS PRODUCT HAS BEEN CALIBRATED IN THE FACTORY PLEASE REFER TO THE MANUAL BEFORE RE CALIBRATING PRODUCT Use...

Страница 2: ...troduction 4 Safety 5 Weight Scale 6 Functions 7 Features 8 Application Conformity 8 RS232 Data 9 Getting Started 10 Delivery 10 Assembly Installation 11 Working Mode 12 Cautions 12 Application Menu 1...

Страница 3: ...uct to which this declaration refers conforms to the following standards Electronic scale Vet Scales series Imperial version Metric version LWC M 800 LWC M 400 Mark applied EU Directive Standards 2004...

Страница 4: ...ion What you should know about these Operating Instructions Tree Professional Weighing Equipment products are simple to operate Nevertheless you should read through these operating instructions in the...

Страница 5: ...r adapter If there should be any reason to believe that it is no longer safe to operate the scale it should be immediately unplugged from the electricity supply pull out power adapter and secured agai...

Страница 6: ...09 2021 6 Weight Scale Construction Functions The weighing scale consists of the following parts The Scale body 1 The Wireless indicator 2 The adapter 3 Operation manual Figure 2 1 Your weighing mach...

Страница 7: ...rial Specifications Model number Capacity Graduation Platter Size LWC P 800 800lb 0 2 lb 5 8 x 37 4 in Shipping weight 38 5 lbs Package Standard carton 40 2 x 10 3 x 10 3 in Operating Temp 40 95 F Pow...

Страница 8: ...eir application in accordance with instructions must be observed The weighing scale has been manufactured and tested in accordance with the standards and recommendations set out in the declaration of...

Страница 9: ...9600 Data Bit 8 Parity Bit None Stop Bit 1 Data Format ASCII 3 Transmission Information Format 20 Byte blank 20H 1 2 3 4 13 14 18 19 20 W P DATA UNIT CR LF W Start of Data Transmission 57H 3AH P Polar...

Страница 10: ...d for a couple hours and kept in its box in a dry room at normal temperature so that no condensation settles on the unit when opening the box Check the scale immediately after unpacking for any extern...

Страница 11: ...age If the operating voltage is not the same as the mains voltage the power adapter must not be connected to the mains socket and contact customer service Placement of Scale The location in which the...

Страница 12: ...ay the self test character of PL or Pr The reason for this phenomenon may be that the power switch of the left scale body or right scale body isn t turned on Another possibility is that the battery of...

Страница 13: ...he ON OFF key is to turn the Scale On or Off The HOLD key is to store the weight value The UNIT key is to toggle different weighing units The ZERO key Zero feature if load weight below 4 of full capac...

Страница 14: ...to turn on the scale the display will show A ON or A OFF indicating that the scale enters the menu of setting auto shut off Press ZERO key to select ON or OFF a ON means auto shut off function activat...

Страница 15: ...ction mode Bluetooth reconnect When reconnect first reconnect the left scale and then reconnect the right scale 1 Turn the switch of the left scale on 2 Turn the switch of the right scale off 3 Press...

Страница 16: ...s the ON OFF key to turn on the indicator the display will show CAL L 4 Press the HOLD key the display will show L0XXX calibration weight press the UNIT key to select unit of measurement kg or lb pres...

Страница 17: ...er plug from socket Also ensure that the scale cannot be connected to the power supply during the work by a third party Make sure that no liquid spills into the scale while performing maintenance work...

Страница 18: ...r purchase After 30 days expire the customer is responsible for shipping the product back to us After the product is received we will inspect it and as necessary we will repair or replace and will shi...

Отзывы: