![luxvision ET-185 Скачать руководство пользователя страница 5](http://html.mh-extra.com/html/luxvision/et-185/et-185_user-manual_1952868005.webp)
4
!
Clean the plastic parts such as chin-rest bracket, forehead-rest belt with soft cloth
dipped with soluble detergent or water, sterilize with medicinal alcohol.
Replacing the fuse
!
Turn off the main power switch and pull out the power plug from the power socket.
!
Screw off the fuse holder cover with the screw driver.
!
Replace with a new fuse, then fasten the cover.
!
The fuse specifications are as follows:
110V
5A/125V;
220V
3A/250V
5 ACCESSORIES
!
Standard table
(
400mmX600mm
)
6 SPECIFICATIONS
Dimensions
:
510mm
(
W
)
X450mm
(
D
)
Table minimum height/stroke
:
660mm/220mm
Weight
:
23Kg
(
body only
)
Maximum load
:
50Kg
Input power supply
(
220Vsystem
)
:
AC220V,230V,240V
Input power supply
(
110Vsystem
)
:
AC110V,120V
Input power supply
(
switchable system
)
:
AC110V,120V,220V,230V,240V
Input power supply
(
system with foot switch
)
:
AC220V,230V,240V
Input power frequency
:
50/60H
Z
Power consumption
:
220VA
Electric safe standard
:
Conform to Standard IEC601 -1
Class I Type B
* Design and specifications are subject to change without notice.