TROUBLESHOOTING
Problem Cause
Action
No light output
Both switches are not “on”
Power not applied to the unit
Attenuator closed
Blown fuse
Turn main power switch on/off and on/
standby switch “on”
Ensure the power cord is connected to
a live, hospital-grade, receptacle
Rotate attenuator clockwise
Replace fuse as indicated in
maintenance section
No light output and unit
is making clicking
sound
Lamp is too old or is defective
A lamp is not installed
Replace/Install lamp. See lamp
replacement section of this user manual
Reduced light output
Attenuator mis-positioned
Aged or defective lamp
Headlight or fiber optic light
guide is damaged
Rotate attenuator
Replace lamp
Check or replace faulty component
Unit feels warm when
not on
Main power switch is on (Figure
3)
This is normal when main power switch
is on and stand by switch is off
CLX enclosure is
overheated
Possible fan failure
Return the unit (See Return and Repair
section of this user manual)
MAINTENANCE AND CLEANING
Maintenance:
If the Light Source does not operate properly when connected to a grounded receptacle, disconnect
the power cord and check the fuse (see below). Do not attempt to repair the unit if the lamp fails in
use. Turn off the unit and allow it to cool for at least 15 minutes, then try to restart the unit. If the lamp
still fails to illuminate, it may need to be replaced. If you still experience difficulties, return the unit to
Luxtec or an authorized distributor for evaluation.
To replace the fuse:
The fuses for the Light Source are located in the power entry module in the rear of the unit. Remove
the power cord from the unit. Using a small flat head screwdriver, open fuse access cover then pry out
the red plastic fuse holder from the power entry module. Check to see if the fuses are blown. If
necessary, replace the fuse(s) with a new one of the same rating (Fuse 3.15A, 250VAC, 5x20 mm,
Slo-Blo, IEC Standard, Luxtec part number 601919). After replacing fuse(s), push the red plastic fuse
holder back into the power entry module and snap access cover close. Plug the cord back into Light
Source and re-test the unit.
Cleaning:
Allow unit to cool for at least 15 minutes prior to cleaning.
The Light Source exterior can be cleaned and disinfected using 70% isopropyl alcohol. Unplug the
power cord before cleaning. Allow 5 minutes for alcohol to evaporate before reconnecting to power.
Use a vacuum cleaner and a soft brush to remove visible dust accumulation from fan and vent holes
whenever necessary and always when replacing the lamp.
- 9 -