19
Ethernet TCP Configuration:
1.
IP Address will be 192.168.0.1 by factory default.
2.
Subnet will be 255.255.255.0 by factory default.
3.
Port is 502 by factory default.
The data transfer is represented on the right-hand side of the screen as TX (data send)
and RX (data receive) messages, accompanied by their time-stamps. This is the raw data
sent and received from the software to your BeST-SLED® unit, and is used to monitor
communications.
Logging:
The Best-SLED® ISB2 supports continuous logging of data such as the currents,
temperatures, fan flow, etc. by clicking on the
“
Logging
”
tab from the main console.
The window shown in Figure 14 will open. Enter the logging file name and directory for
where the data is to be stored by
clicking on the “…” button
. After the filename and
path have been selected, to start logging, click the check box before
“
Log to File
”
to
enable the logging.
Note: Log Files must use the .csv (Comma Separated Value) file extension.
Figure 14: Data Logging Window