Lutron Electronics DS-9200 Скачать руководство пользователя страница 4

2. SPECIFICATIONS

Circuit

&XVWRPVLQJOHFKLSPLFURSURFHVVRU

 LSI

circuit.

Display

LCD size : 35 x 30 mm.
LCD with green backlight ( ON/OFF ).

Measurement 

SPL 

: Sound pressure level

Type

Dose

test noise exposure

Measurement 

SPL

Auto 35 - 130 dB.

Range

Dose

70

130dB

Resolution

0.1 dB.

Function

dB ( A & C frequency weighting ), 
Time weighting ( Fast, Slow ), 
Data hold
Record ( Max., Min. ).

Accuracy

Characteristics of " A " frequency weighting

(23 ±5 

)

network meet IEC 61672 class 2.
Under 94 dB input signal,  the accuracy
are :

 31.5

Hz

 ±3.0 dB

 63

Hz

 ±2.0 dB

 125

Hz

 ±1.5 dB

 250

Hz

 ±1.5 dB

 500

Hz

 ±1.5 dB

 1 K

Hz

 ±1.0 dB

 2 K

Hz

 ±2.0 dB

 4 K

Hz

 ±3.0 dB

 8  K

Hz

 ±5.0 dB

Remark :
The above spec. are tested under the
environment  RF Field Strength less
than 3 V/M & frequency less than 30
MHz only.

2

Содержание DS-9200

Страница 1: ...sonal sound exposure Noise Dosimeter with DATALOGGER marks a step forward for you into the field of precision measurement Although this METER is a complex and delicate instrument its durable structure...

Страница 2: ...sampling time 1 second 11 5 3 Manual Datalogger Set sampling time 0 second 13 5 4 Check time and sampling time information 13 6 ADVANCED SETTING 14 6 1 Set clock time Year Month Date Hour Minute Seco...

Страница 3: ...WHULVWLF PRGHV 8VHU GHILQHG PHDVXUHPHQW FRQILJXUDWLRQV 6HOHFWDEOH ULWHULRQ HYHO FKDQJH 5DWH 7KUHVKROG UHTXHQF HLJKWLQJ DQG 5HVSRQVH 7LPH VHWWLQJ RQGHQVHU PLFURSKRQH IRU KLJK DFFXUDF ORQJ WHUP VWDELOLW...

Страница 4: ...B Function dB A C frequency weighting Time weighting Fast Slow Data hold Record Max Min Accuracy Characteristics of A frequency weighting 23 5 network meet IEC 61672 class 2 Under 94 dB input signal t...

Страница 5: ...check knowing the sound pressure level of the tested equipment Time Fast t 125 ms weighting Fast range is simulating the human ear Fast Slow response time weighting Slow t 1 s Slow range is easy to g...

Страница 6: ...r Month Date Hour Minute Second Set sampling time Auto power OFF management Set Criterion level in dB Set Threshold in dB Set Exchange rate in dB Check Memory space Set Clear Memory Overage Indication...

Страница 7: ...DC 8 mA AC DC power adapter is optional If LCD backlight on the power consumption will increase approx DC 10 mA Weight 250 g 0 55 LB without battery Dimension HWD 116 x 45 x 20 mm 4 6 x 1 8 x 1 2 inch...

Страница 8: ...splay 3 5 Hanging holes 3 10 Battery compartment and Screw of the battery cover 3 11 Tripod Fix Nut 3 12 RS232 Output 3 13 Microphone input jack 3 14 Microphone 3 6 Button DUR BEG Fast Slow 3 15 Micro...

Страница 9: ...tening response Typically always select the A weighting when makes environmental sound level measurement makes environmental sound level measurement b The C weighting characteristic is near the FLAT c...

Страница 10: ...ng mode if 8 sec without chang seting will ESC the seting mode Note a LC Criterion Level 80 84 85 90dB b LT Threshold Level 70 to 90dB in 1 dB steps c ER Exchange Rate 3 4 5 or 6 db d pressing the Pow...

Страница 11: ...d dose every day Please use the button to select R BEG then press RUN button If user wants to rerun dose Function Please use the button and button 2 sec to reset dose value and time If user wants to c...

Страница 12: ...appear on the display Press the HOLD Button again will show the REC symbol only execute the memory function continuously b To exit the memory record function just press the RUN button 2 seconds at le...

Отзывы: