background image

  micro SD card real time data logger
  CAT III 1000V, true rms, L/C

MULTIMETER

  Model : DM-9962SD

Your purchase of this
MULTIMETER marks a
step forward for you
into the field of
precision measurement.
Although this
MULTIMETER is a
complex and delicate
instrument, its durable
structure will allow
many years of use if
proper operating
techniques are
developed. Please read
the following
instructions carefully
and always keep this
manual within easy
reach.

OPERATION MANUAL

OPERATION MANUAL

Содержание DM-9962SD

Страница 1: ...ward for you into the field of precision measurement Although this MULTIMETER is a complex and delicate instrument its durable structure will allow many years of use if proper operating techniques are...

Страница 2: ...oltage current to the input terminal Remove test leads before open the battery cover Cleaning Only use the dry cloth to clean the plastic case Environment Conditions Installation Categories III 1000V...

Страница 3: ...asurement 10 5 5 Continuity Check 11 5 6 Diode Test 11 5 7 Capacitance Measurement 13 5 8 Inductance Measurement 13 5 9 L C Calibration 14 5 10 Frequency Measurement 14 5 11 Temperature Measurement 15...

Страница 4: ...OFF management 24 8 4 Set beeper sound ON OFF 25 8 5 Set Decimal point of Micro SD card setting 25 8 6 Set clock time Year Month Date Hour Minute Second 26 8 7 Set Change the Temp unit 26 9 MAINTENAN...

Страница 5: ...istance Capacitance Inductance Frequency Temperature Diode Continuity beeper Max Min measurement value with recall Relative Data hold Auto range with manual range selection Temperature measurement pos...

Страница 6: ...eeze the display reading on the LCD display Power On Off Auto power off or manual power off management Details please refer page 7 Memory recall Records Maximum Minimum readings with recall Relative m...

Страница 7: ...Sub 9 connector UPCB 06 RS232 cable to USB connector USB 11 Micro SD card AC MICROSD Data Acquisition software SW U801 WIN SW E802 2 2 Electrical Specifications 23 5 DC Voltage Range 600 0 mV 6 V 60...

Страница 8: ...age VF Range 2 9 V DC Accuracy 0 5 2d Capacitance Capacitance Capacitance Capacitance Auto Range 600 pF 6 nF 60 nF 600 nF 6 uF 60 uF 100 uF 600 uF Resolution 0 1 pF 0 001 nF 0 01nF 0 1 nF 0 001 uF 0 0...

Страница 9: ...M 1 5 2d 60 M 3 5d Over load 350 DCV 350 ACV protection Continuity Beeper Beeper will sound if measured resistance less than 20 ohm Max Min Measurement During the operation can memorize the maximum a...

Страница 10: ...3 3 TIME SET button 3 12 COM input terminal 3 4 REL Backlight button 3 13 mA uA input terminal 3 5 HOLD button 3 14 10A input terminal 3 6 RANGE button 3 15 Battery compartment Cover 3 7 V A Hz button...

Страница 11: ...e Function Rotary Switch to the Off position when the instrument is not operation 7 Remove the battery if the instrument is not to be used in a long period of time 8 For safety consideration when chan...

Страница 12: ...ax Appears when Max and Min value record Min function is operated Battery voltage is already under the low condition Appears when the Continuity beeper is operated mV V Units for voltage measurements...

Страница 13: ...ll execute the Manual Range mode and hold the range the LCD will show the MANU marker Under the manual range operation push the Range button 3 6 Fig 1 2 seconds will return to auto range operation Rem...

Страница 14: ...Fig 1 For the 10 A current measurement connect RED test lead into A terminal 3 14 Fig 1 Open the circuit in which current is to be measured Now securely connect test leads in series with the load in...

Страница 15: ...al 2 Connect RED test lead into terminal 3 Select the Function rotary switch 3 10 Fig 1 to the position 4 The LCD display will show the marker 5 when the resistance value is less than 20 ohm the beepe...

Страница 16: ...3 a reverse check on the diode is made If the diode under test is good will be displayed If the diode under test is defective 0 000 or other numbers will be displayed Proper diode testing should incl...

Страница 17: ...When LCD show the AUTO marker the meter is under the auto range mode Meter will select the suitable measurement range automatically 4 Capacitance measurement mode no manual functions Range button 3 6...

Страница 18: ...calibration data is saved to the EEPROM complete the calibration 5 10 Frequency Measurement 1 Connect BLACK test lead into COM terminal 3 12 Fig 1 2 Connect RED test lead into Hz terminal 3 11 Fig 1...

Страница 19: ...3 4 Fig 1 at once then LCD will show zero value a REL indicator 2 The input measured values will deduct last measured values automatically then show those new value on the display 3 It will cancel th...

Страница 20: ...n the display will show the MIN maker along with the minimum measured value 4 If intend to cancel the MAX MIN Record function just push the MAX MIN button 3 2 Fig 1 2 seconds the REC marker will disap...

Страница 21: ...Hz 31 DC A 36 AC A 52 KHz 33 38 pF A4 MHz 67 K 39 uF 44 01 M 40 nF 43 02 DIODE 46 D10 Polarity 0 Positive 1 Negative D9 Decimal Point DP position from right to the left 0 No DP 1 1 DP 2 2 DP 3 3 DP D8...

Страница 22: ...ing format by the meter If the micro SD card exist the trouble during format by the meter If the micro SD card exist the trouble during format by the meter If the micro SD card exist the trouble durin...

Страница 23: ...again the LOGGER symbol will flashing c Finish the Datalogger During execute the Datalogger function press the LOG Button 3 8 Fig 1 1 5 seconds continuously again will finish the Datalogger function...

Страница 24: ...time and sampling time information During the normal measurement not execute the Datalogger If press Time 3 3 Fig 1 once the LCD display will present the time information of Year Month Date Hour Minu...

Страница 25: ...d socket 3 18 Fig 1 2 Plug in the Micro SD card into the Computer s Micro SD card slot if your computer build in this installation or insert the Micro SD card into the Micro SD card adapter then conne...

Страница 26: ...XCEL data screen for example EXCEL graphic screen for example graphic 22 50 70 90 110 130 150 170 190 210 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 A C V Graphic Ch1_Valu...

Страница 27: ...OFF management bEEP Set beeper sound ON OFF dEC Set Micro SD card Decimal character dAtE Set clock time Year Month Date Hour Minute Second t CF Set Temperature unit or Remark In the implementation of...

Страница 28: ...nual Datalogger or 2 3600 seconds to Auto Datalogger the adjusted value will be flashed When the desired value is set press the Enter key 3 8 Fig 1 once save the default sample value 2 sp t setting is...

Страница 29: ...unction with default 8 5 Set Decimal point of Micro SD card setting When the display show dEC The numerical data structure of Micro SD card is default used the as the decimal for example 20 6 1000 53...

Страница 30: ...rst setting value is Year then next to adjust Month Date Hour Minute Second value Remark The adjusted value will be flashed 2 After set all the time value Year Month Date Hour Minute Second press the...

Страница 31: ...attery cover opening the battery cover opening the battery cover 1 When the LCD display showing the mark of it is necessary to replace the battery However in spec measurement may still be made for sev...

Страница 32: ...not operation please check if the Fuse A is broken or not When the 10 A current range can not operation please check if the Fuse B is broken or not 3 When replace the fuse should take the test leads f...

Страница 33: ...ect the meter to the computer COM port USB cable Computer interface cable USB 11 Used to connect the meter to the computer USB port Item Model Carrying Case CA 05A Light Adapter LX 02 EMF Adapter EMF...

Страница 34: ...11 THE ADDRESS OF AFTER SERVICE CENTER 30 180525 DM 9962SD...