background image

Replacing the Flow Cell

When replacing the flow cell, handle it with care. The capillary is 
fragile; avoid touching it unnecessarily. Do not force the flow cell 
into the receptacle.

1.

Remove the metal plate from the top of the instrument.

2.

Remove the minstac tubing from the clamp and disconnect it 
from the instrument.

3.

Remove the flow cell by grasping it at the top with your fingers 
and pulling straight up. Do not pull up on the tubing. You can 
also use the flow cell removal tool to remove the flow cell.

•  

Grasp top of the flow cell with the removal tool by placing 

the clamps into the cutouts at the top of the flow cell, as 
shown.

•  

Pull the tool straight up to remove the flow cell.

4.

Install the new flow cell by correctly positioning it above the 
instrument and carefully lowering it into the receptacle. It fits 
one way only—note the beveled edge. 
Avoid bumping the glass capillary against the instrument or 
sides of the receptacle.
Push down firmly on the top of the flow cell on either side of the 

tubing using your fingers until the flow cell clicks into place. Do 
not press down on the tubing.

5.

Connect the tubing to the instrument, then insert the tubing 
into the clamp.

6.

Run Guava® Clean to prime the instrument, followed by easy-
Check™.

Technical Support

Telephone: 512-381-4397
North America Toll Free: 1-877-785-2323
International Toll Free: + 800-2939-4959
Email: 

[email protected]

www.luminexcorp.com

Guava is a trademark of Luminex Corporation, registered in the U.S. and other countries. 
easyCheck and easyCyte are trademarks of Luminex Corporation.
Part No. 0110-8490 Rev D, July 2019, Printed in the USA.
© 2014 - 2019 Luminex Corporation. All rights reserved.

Отзывы: