Application Note 4 Reliable Operation of LSProbe 1.2 Field Probes

nectors at the other end. The two sacrificial cables are connected by an E2000 coupler.
The FC/ST connectors at either end of the sacrificial cable kit connect to and protect the
FC/ST connectors of the Field Probe, the Computer Interface and the fiber extension cable.
Fig. shows the fiber connections of a typical LSProbe 1.2 Field Probe System.

Figure 5: Typical LSProbe 1.2 Field Probe System with sacrificial cable kits

As highlighted in Fig. 6, the optical link should always be disconnected using E2000 connec-
tors. Do not disconnect any FC/ST connectors unless absolutely necessary. When plugging
FC/ST connectors make sure to observe the cleaning instructions in Section 4.

F

C

/S

T

 

C

o

u

p

le

r

Field

Probe

Use E2000 

Connectors!

Two Sacrificial Cable Kits incl. 

E2000 Connectors and Coupler 

E2000

Fiber

Coupler

FC/ST

Fiber

Coupler

FC/ST

Fiber

Coupler

E2000

Fiber

Coupler

Fiber-Optic Cable

Do Not 

Disconnect¹!

Fiber-Optic 
Extension 
Cable

typ. 5m

typ. 15m

1) Except when replacing the sacrificial cables.

Do Not 

Disconnect¹!

FC/ST

Fiber

Coupler

Computer
Interface

FC/ST

Fiber

Coupler

typ. 20cm

Figure 6: Fiber connections of a typical LSProbe 1.2 Field Probe System

November 23, 2020

Page 5/9

Содержание LSProbe 1.2

Страница 1: ... Probes LUMILOOP GmbH support lumiloop de November 23 2020 Contents 1 Introduction 2 2 Fiber Failure Mechanism 3 3 Sacrificial Cable Kits for LSProbe 1 2 Field Probe Systems 4 4 Fiber Connector Cleaning Instructions 6 5 Replacing Sacrificial Cables 8 6 Fiber Handling Recommendations 9 ...

Страница 2: ...ectors of Field Probe Computer Interface and extension cable They provide two defined and convenient points for disconnecting the Field Probe from the Computer Interface using the E2000 con nectors of the sacrificial cable kits Due to the integrated shutters burn in defects at the E2000 connectors are significantly less frequent However in the case of a burn in defect at an E2000 connector the sacrifi...

Страница 3: ...C fiber connector pair which carries the optical power for the field probe suffered a burn in defect a Field Probe FC before after cleaning b Field Probe ST before after cleaning c Extension cable FC before after cleaning d Extension cable ST before after cleaning Figure 2 Example for contaminated FC power and ST data fiber connectors As suggested by the name burn in defects are caused by excessive he...

Страница 4: ...tors are significantly less prone to contamination and thus to burn in defects because of their integrated automatic shutters Consequently E2000 connectors may be plugged without cleaning consciously accepting that contamination will accumulate with each plugging cycle and lead to a burn in defect eventually Once a burn in defect has occurred the complete sacrificial cable kit must be replaced in cl...

Страница 5: ...al link should always be disconnected using E2000 connec tors Do not disconnect any FC ST connectors unless absolutely necessary When plugging FC ST connectors make sure to observe the cleaning instructions in Section 4 FC ST Coupler Field Probe Use E2000 Connectors Two Sacrificial Cable Kits incl E2000 Connectors and Coupler E2000 Fiber Coupler FC ST Fiber Coupler FC ST Fiber Coupler E2000 Fiber ...

Страница 6: ...n inad vertently during fiber inspection Note that the focal plane for for E2000 and ST FC connectors differ and must thus be adjusted 2 Clean the connector end face 3 Reinspect the connector end face go back to step two if there is any contamination 4 Connect the fiber connectors do not put the connector down Do not apply dust caps between cleaning and connecting LUMILOOP s cleaning kit shown in Fig...

Страница 7: ...rm Step 3 Moisten one corner of the cleaning wipe us ing the fiber optic cleaning pen or IPA pipette Step 4 Press the fiber end face flatly against the moist area of the wipe and slide it into the dry area of the wipe Step 5 Only for E2000 connectors remove the E2000 locking cap re clip the connector pair after cleaning both connectors Figure 8 FC ST E2000 fiber optic connector wet to dry cleaning met...

Страница 8: ...or and connect it to the coupler d Repeat steps a through c for the ST connectors 4 Repeat step three for the second end of the sacrificial cable kit 5 Consider ordering a new spare sacrificial cable kit Fixing burned in FC ST connectors In case of a burn in defect at a ST or FC connector due to improper handling replacement is required There are three destinct cases In case of a burn in defect at a...

Страница 9: ...g Avoid tensile strain on fiber optic cables Fiber optic cables shall not carry the weight of the connectors Always use two sacrificial cable kits one between Computer Interface and extension cable and one between extension cable and Field Probe Always disconnect the optical fiber connection using the E2000 connectors of the sacrificial cable kits Do not open any FC ST fiber connectors unless absolutel...

Отзывы: