14.0 Troubleshooting
SYMPTOM:
CORRECTIVE ACTION
Mattress does not
inflate or is too soft.
Make certain the air hoses are not kinked, cracked, or split. Verify the hose couplings
are fully engaged to controller. Verify that all 20 air cells are connected to the air manifold
within mattress.
Power loss indicator
flashing and audible
alarm sounding
Power has been disconnected from unit while in an operational mode.
Restore power, put unit in STANDBY mode, then remove power. Unit must always be
returned to STANDBY mode before disconnecting power.
Indicator adjacent to
ALARM SILENCE
button is flashing and
audible alarm is
sounding.
If in AUTO mode, check the connection of the Active Sensor mattress sensor cord to the
jack on the side of the unit. Reconnect if necessary.
Verify hose couplings are fully engaged to controller. Verify that all 20 air cells are
connected to the air manifold within mattress.
15.0 Specifications, Control Unit
Enclosure Dimensions
10" x 12" x 4
1
/
2
", (25 cm x 30 cm x 11 cm)
Weight
9.5 pounds (4.3 kg)
Power Cord
Detachable 14' , #18 AWG minimum, with ground wire
Overcurrent Protection
Primary 2.5A, 250V, T
Input
120 -230 VAC, 50/60Hz, 120V – 0.44A, 230V – 0.33A
Operating Ambient
Temperature Range
60 to 85°F (16 to 29°C)
Classification
Class I grounded equipment not suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide.
Type BF equipment
UL2601-1, CSA C22.2 NO. 601.1
EN 60601-1
IPXO, enclosed equipment without protection
against ingress of water.
Continuous operation
Electromagnetic
Compatibility
Meets EN60601-1-2:2001
(CISPR 11 Classified as Class B, Group 1 ISM equipment)