Troubleshooting/Specifications

14.0 Troubleshooting

SYMPTOM:

CORRECTIVE ACTION

Mattress does not

inflate or is too soft.

Make certain the air hoses are not kinked, cracked, or split. Verify the hose couplings 

are fully engaged to controller. Verify that all 20 air cells are connected to the air manifold

within mattress.

Power loss indicator

flashing and audible

alarm sounding

Power has been disconnected from unit while in an operational mode. 

Restore power, put unit in STANDBY mode, then remove power. Unit must always be

returned to STANDBY mode before disconnecting power.

Indicator adjacent to

ALARM SILENCE

button is flashing and

audible alarm is

sounding.

If in AUTO mode, check the connection of the Active Sensor mattress sensor cord to the

jack on the side of the unit. Reconnect if necessary.

Verify hose couplings are fully engaged to controller. Verify that all 20 air cells are

connected to the air manifold within mattress.

15.0 Specifications, Control Unit

Enclosure Dimensions

10" x 12" x 4 

1

/

2

", (25 cm x 30 cm x 11 cm)

Weight

9.5 pounds (4.3 kg)

Power Cord

Detachable 14' , #18 AWG minimum, with ground wire

Overcurrent Protection

Primary 2.5A, 250V, T

Input

120 -230 VAC, 50/60Hz, 120V – 0.44A, 230V – 0.33A

Operating Ambient
Temperature Range

60 to 85°F (16 to 29°C)

Classification

Class I grounded equipment not suitable for use in the presence of a flammable

anesthetic mixture with air or with oxygen or nitrous oxide.

Type BF equipment

UL2601-1, CSA C22.2 NO. 601.1

EN 60601-1

IPXO, enclosed equipment without protection 

against ingress of water.

Continuous operation

Electromagnetic
Compatibility

Meets EN60601-1-2:2001

(CISPR 11 Classified as Class B, Group 1 ISM equipment)

Содержание ALTADYNE 753003

Страница 1: ...05 AltaDyne AutoFloat Alternating Pressure Management Plus Model 753003 Model 753004 OPERATORS MANUAL Graham Field Health Products 2935 Northeast Parkway Atlanta Georgia 30360 1 800 347 5678 Fax 1 80...

Страница 2: ...ded contact the Graham Field Health Products technical service department Domestic Customers Toll Free 1 800 347 5678 Fax 1 800 726 0601 1 0 Warranty The Control Unit is warranted free of defects in m...

Страница 3: ...d frame to prevent mattress from sliding and causing patient injury Lumex mattresses are not intended to be AND DO NOT FUNCTION AS a patient fall safety device SIDE RAILS MUST BE USED WITH THE LUMEX M...

Страница 4: ...than that of the patient s skin When the patient moves the Top Sheet tends to move relative to the air cells rather than relative to the patient minimizing shearing effects Additionally the quilting...

Страница 5: ...l light while AP mode is active Note AP mode is temporarily suspended in max inflate mode 4 SOFT decreases the inflation level of the bed while Control Unit is in manual mode only Pressing once will d...

Страница 6: ...ct the dual quick connect couplings of the mattress hoses to the mating connectors located on the side of the Control Unit 3 Check that the hoses are secure by gently pulling Ensure the hoses are not...

Страница 7: ...y before resuming the A P operation 10 2 Use With A Static Mattress 1 Press the Mode Select button until the icons adjacent to the MANUAL and FLOAT icons are both illuminated 2 Adjust the inflation le...

Страница 8: ...f mattress become soiled clean and disinfect as described above or simply replace air cell with a clean replacement Single air cell replacement can be successfully achieved with patient remaining on t...

Страница 9: ...ttress sensor cord to the jack on the side of the unit Reconnect if necessary Verify hose couplings are fully engaged to controller Verify that all 20 air cells are connected to the air manifold withi...

Отзывы: