AltaDyne AutoFloat

Description/Features

6.0 Description

The AltaDyne AutoFloat is a portable air flotation system with on-demand alternating low pressure treatment,

designed to provide benefit to patients suffering from, or at risk of, developing pressure ulcers. The system,

consisting of a Control Unit and Air Flotation Mattress with it’s Top Sheet, is designed to provide pressure

management and patient comfort.

The Control Unit inflates and maintains the Mattress to user-selected settings, and can also provide on demand

alternating low pressure treatment if desired. Featuring Active Sensor Technology, the system automatically adjusts

the Mattress to the optimal fill level ensuring complete flotation pressure management.

Mattress

The Mattress base consists of twenty transverse air cells constructed of a low shear nylon. Each of the individual

cells is a minimum of eight inches tall (inflated height) and spatially oriented above a two inch convoluted foam base.

The Active Sensor Mattress utilizes 4 specialized air cells in the center region of the Mattress which contain active

sensors for detecting patient flotation level.

Covering the entire Mattress assembly is a low-friction, low-shear producing, vapor permeable, nylon Top Sheet. 

The Top Sheet surface creates a membrane impermeable to liquids, air and bacteria though still permeable to water

vapor. Quilted onto the base of the Mattress Top Sheet is a spun bonded polyester fiberfill that exhibits less frictional

resistance to nylon than that of the patient’s skin. When the patient moves, the Top Sheet tends to move relative to

the air cells rather than relative to the patient, minimizing shearing effects. Additionally, the quilting acts to provide 

a diffusion layer by which the water vapor can be drawn away from the patient, resulting in more efficient

evaporation and subsequent reduced skin maceration.

7.0 Features

7.1 Control Unit

• Max inflate for easy patient transfer.

• Soft/firm comfort control adjustment for customized treatment.

• On-demand, alternating, low pressure management.

• Automatic patient flotation control via Active Sensor Technology™ to ensure optimal pressure management.

• CPR quick release providing for rapid deflation.

• Sleek and durable plastic housing is extremely compact and lightweight.

• Non-slip mounting brackets for secure placement on the foot board of bed.

• Illuminated LED control panel for easy viewing.

• Control panel Lockout Mode to prevent inadvertent patient setting changes.

• Alarm Silence feature to temporarily suspend audible alarm.

7.2 Active Sensor Air Flotation Mattress

• Horizontal, tubular cell construction for optimal comfort and pressure management.

• Individual air cells designed for maximum support and pressure distribution.

• High density foam base for additional protection.

• Low friction, low shear Top Sheet to aid in prevention of skin breakdown.

• Dimensions - base model 10" x 34" x 80" (25 cm x 86 cm x 203 cm) when fully inflated. Other sizes are available.

• The red air cells contain active sensors for automatic patient flotation control.

• Optional Footstop cell to minimize heel interface pressure while maintaining proximal positioning of feet.

• Removable cells for greater flexibility and easy maintenance.

Содержание ALTADYNE 753003

Страница 1: ...05 AltaDyne AutoFloat Alternating Pressure Management Plus Model 753003 Model 753004 OPERATORS MANUAL Graham Field Health Products 2935 Northeast Parkway Atlanta Georgia 30360 1 800 347 5678 Fax 1 80...

Страница 2: ...ded contact the Graham Field Health Products technical service department Domestic Customers Toll Free 1 800 347 5678 Fax 1 800 726 0601 1 0 Warranty The Control Unit is warranted free of defects in m...

Страница 3: ...d frame to prevent mattress from sliding and causing patient injury Lumex mattresses are not intended to be AND DO NOT FUNCTION AS a patient fall safety device SIDE RAILS MUST BE USED WITH THE LUMEX M...

Страница 4: ...than that of the patient s skin When the patient moves the Top Sheet tends to move relative to the air cells rather than relative to the patient minimizing shearing effects Additionally the quilting...

Страница 5: ...l light while AP mode is active Note AP mode is temporarily suspended in max inflate mode 4 SOFT decreases the inflation level of the bed while Control Unit is in manual mode only Pressing once will d...

Страница 6: ...ct the dual quick connect couplings of the mattress hoses to the mating connectors located on the side of the Control Unit 3 Check that the hoses are secure by gently pulling Ensure the hoses are not...

Страница 7: ...y before resuming the A P operation 10 2 Use With A Static Mattress 1 Press the Mode Select button until the icons adjacent to the MANUAL and FLOAT icons are both illuminated 2 Adjust the inflation le...

Страница 8: ...f mattress become soiled clean and disinfect as described above or simply replace air cell with a clean replacement Single air cell replacement can be successfully achieved with patient remaining on t...

Страница 9: ...ttress sensor cord to the jack on the side of the unit Reconnect if necessary Verify hose couplings are fully engaged to controller Verify that all 20 air cells are connected to the air manifold withi...