Test voltage:
2210 V AC rms 1 minute between housing / power supply and:
- RS485
- binary outputs
- analog inputs
1390 V AC rms 1 minute between:
- analog inputs / RS485
- analog inputs / binary outputs
- RS485 / binary outputs
Power consumption:
≤
13 VA
Weight
< 0.4 kg
Dimensions
48 X 144 X 100 mm
EMC compatibility:
- immunity to interference in accordance with EN 61000-6-2
- interference emission in accordance with EN 61000-6-4
Safety requirements:
in accordance with the standard EN 61010-1
•
insulation between circuits
basic
•
installation category
III,
•
degree of pollution
2,
•
maximum voltage relative to earth:
- for power circuit
300 V
- for input circuit
600 V
- for other circuits
50 V
•
altitude ASL< 2000 m