Lumax LX-1371 Скачать руководство пользователя страница 4

The Lumax Electronic Flow Meter features a menu with which the user can select as the main unit measurement: Quarts (Qts), Pints (Pts), Litres (Lit) or Gallons (Gal). The combination of the unit of 

measurement of the Partial register and that of the Totals is predefined according to the following table:

 

Combination  

Unit of Measurement  

Register Unit of Measurement

 

No. 

of the Partial Register 

of the Totals Register

 

1  

Litres (L)  

Litres (L)

 

2  

Gallons (Gal)  

Gallons (Gal)

 

3  

Quarts (Qts)  

Gallons (Gal)

 

4  

Pints (Pts)  

Gallons (Gal)

Maintenance:

 The Lumax Electronic Flow Meter has been designed to require a 

minimum amount of maintenance. The only maintenance required are the following:

  1.   Battery Change

  2.  Cleaning of the Turbine

1.  Battery Replacement:

  The Lumax Electronic Flow Meter comes complete with 

  2 x 1.5 V. Alkaline Batteries (Size AAA). Lumax Electronic Flow Meter features 

  two low-battery alarm levels:

a.

  When the battery charge falls below the first level on the LCD, the fixed battery symbol appears. In this condition, 

  Flow Meter continues to operate correctly, but the fixed icon warns the user that it is advisable to change the batteries.

b.

  If Flow Meter operation continues without changing the batteries, the second battery alarm level will be reached 

  which will prevent operation. In this condition the battery icon starts to flash and is the only icon that remains 

  visible on the LCD. To change the Batteries, with reference to the exploded diagram positions, proceed as follows:

•   Press RESET to update all the totals

•   Loosen the 4 screws on the back cover

•   Remove the old Batteries

•   Place the new batteries in the same position as the old ones

•   Close the cover again, by positioning the rubber protection as a gasket

•  Flow Meter will switch on automatically and normal operation can be resumed.

The Flow Meter will display the same Reset Total, the same Total and the same Partial indicated before the Batteries were changed. 

After changing the Batteries, the meter does not need calibrating again.

Malfunctions

4.  Direct modification of K Factor: 

This procedure is especially useful to correct a “mean error” obtainable on the basis of several performed dispensing operations. If normal Flow Meter operation 

shows a mean percentage error, this can be corrected by applying to the currently used calibration factor a correction of the same percentage. In this case, the percentage correction of the “USER” K 

Factor must be calculated by the operator in the following way:

New K Factor = Old K Factor x    (100 - E%)
                                                          100

If the meter indicates less than the real dispensed value (negative error) the new calibration factor must be higher than the 

old one as shown in the example. The opposite applies if the meter shows more than the real dispensed value (positive error).

(

(

Example:

Error percentage found E%  

- 0.9 %

Current Calibration Factor  

1,000

New “USER” K Factor  

1,000 * [(100 – ( - 0,9))/100]=

 

1,000 * [(100 + 0,9)/100] = 1.009

Action Display

6  SHORT/LONG CAL KEY KEYING

 

The indicated value changes in the direction indicated by the arrow:

 

- one unit for every short “MENU” key keying

 

- continually if the “MENU” key is kept pressed. The speed

 

increase rises by keeping the key pressed. If the desired 

 

value is exceeded, repeat the operations from point (5).

7  LONG RESET KEY KEYING

 

Flow Meter is informed that the calibration procedure is 

 

finished. Before performing this operation, make sure 

 

the indicated value is that required.

8  NO OPERATION

 

At the end of the calculation, the new “USER” K Factor is shown for 

 

a few seconds, after which the restart cycle is repeated to finally 

 

achieve standby condition. 

Attention: 

From now on, the indicated 

 

factor will become the calibration factor used by the meter and will 

 

continue to remain such even after a battery change.

9.  NO OPERATION

 

The Flow Meter stores the new work calibration factor and 

 

is ready to begin dispensing, using the “USER” K Factor 

 

that has just been calculated.

1.003

Q

TS

c

al

DIRECT

-----

Q

TS

c

al

DIRECT

1.003

Q

TS

c

al

end

0.000

Q

TS

13456

TOTAL

G

AL

Action Display

1 NONE

 

Flow Meter in Stand By (not in counting mode)

2  LONG “CAL” KEY KEYING

 

Flow Meter enters calibration mode, shows “CAL” and displays the 

 

Calibration Factor being used instead of the Partial. The words 

 

“Fact” and “USER” indicate which of the two Factors (factory or user) 

 

is currently being used.

3

 

LONG RESET KEY KEYING

 

Flow Meter shows “CAL” and the Partial at zero. Flow Meter is ready 

 

to perform on-site calibration by dispensing.

4

 

LONG RESET KEY KEYING

 

We now go on to Direct change of the calibration factor:The word 

 

“Direct” appears together with the Currently used Calibration Factor. 

 

In the bottom left part of the display, an arrow appears (upwards or 

 

downwards) defining the direction (increase or decrease) of change 

 

of the displayed value when subsequent operations 5 or 6 are performed.

5

 

SHORT RESET KEY KEYING

 

Arrow direction changes. The operation can be 

 

repeated to alternate the direction of the arrow.

12.345

Q

TS

1234.5

TOTAL

G

AL

1.000

(user)

c

al

FACT

12.345

Q

TS

c

al

field

1.000

Q

TS

c

al

DIRECT

1.000

c

al

DIRECT

12.345

Q

TS

23412.3

TOTAL

G

AL

Problem 

LCD: No indication contact

Precision Measurement not enough

Reduced or Zero Flow Rate

The meter does not count, but the Flow Rate is correct

Possible Cause

Bad Battery

Wrong K Factor

The meter works below minimum acceptable Flow Rate

Turbine Blocked

Incorrect Installation of Gears after Cleaning

Possible Electronic Card issues

Corrective Action

Check Battery Contacts

Check the K Factor according to the Instructions

Increase the Flow Rate until an acceptable Flow Rate range has been achieved

Clean the Turbine

Repeat the Reassembly Procedure

Contact your Dealer

WARNING!:

Do not discard the old batteries in the environment. 

Refer to local disposal regulations.

2. Cleaning: 

Only one operation is necessary to clean the Flow Meter. 

After removing Flow Meter from the plant where it was 

built in, any residual elements can be removed by 

washing or mechanically-handling. If this operation 

does not restore a smooth rotation of the turbine, it will 

have to be replaced.

WARNING!: 

Do not use compressed air onto the turbine in order to 

avoid its damage because of an excessive rotation

Form: 

LX-1371-IS

© 2016 Lumax.  All Rights Reserved. 

Lumax® and the Products described in it are Copyrighted  by Lumax®

, with All Rights Reserved. 

This Form may not be Copied in Whole or Part, without the Express Written Consent of Lumax

®.

Viton

® ;  

Ryton

®  and Teflon® are registered trademarks of their respective companies

To choose between the 4 available combinations:

•   Wait for Flow Meter to go to Standby

•   Press the “MENU” and “RESET” keys together and hold for five seconds. The current unit of measure will 

  begin to flash. Press “RESET” to chose a different unit of measure, then press MENU to confirm. 

•   Flow Meter will pass through the start cycle and will then be ready to dispense in the set units.

Note:

   The Reset table Total and Total registers will be automatically changed to the new unit of 

 

measurement. NO new calibration is required after changing the Unit of Measurement.

unit

Q

TS

G

AL

unit

P

TS

G

AL

unit

L

L

unit

G

AL

G

AL

R

R

R

R

Отзывы: