INFRATHERM pyrometer ISR 50-LO 

 

 

37

5.4 

Connecting to RS232 interface 

The transmission rate (in baud) of the serial interface is de-
pendent on the length of the cable

Values between 1200 and 

115200 Bd may be set (the PC program InfraWin accepts 
maximum 38400 Bd). 
The baud rate has to be reduced by 50% when the transmis-
sion distance is doubled (see also 9.9 Baud rate). 
 
Typical cable length for RS232 at 19200 Bd is 7 m. 
 

 

PC’s side (soldering side)

1

 

4

6

 

7

8

9

 DGND (red 

*

)

)

   RxD (black 

*

)

)

   TxD (violet 

*

)

)

*

)

 The color specification refer 

to the standard assignment of 
all IMPAC connection cables 

2

 

3

5

 
5.5 

Connecting to RS485 interface 

The transmission rate (in baud) of the serial interface is dependent on the length of the cable

Values be-

tween 1200 and 115200 Bd may be set (the PC program InfraWin accepts maximum 38400 Bd). 
Typical cable length for RS485 at 19200 Bd is 2 km. The baud rate has to be reduced by 50% when the 
transmission distance is doubled (see also 9.9 Baud rate). 
 
 
5.6 

Connection of additional analyzing devices 

Additional analyzing instruments, for 
example a LED digital display in-
strument only needs to be connected 
to a power supply and the analog 
outputs from the pyrometer. Another 
Instruments like a controller or printer 
can be connected to the display in 
series as shown above (total load of 
resistance max. 500 Ohm). 

230 V ~

 

24 V DC

 

°C

 

yellow 

*

)

 

LED digital display

Controller

Power supply

green 

*

)

 

brown 

*

)

 

white 

*

)

 

Printer

*

)

 The color specification refer to the standard assignment of all IMPAC connection cables 

 
 

6

 

Mechanical Installation 

 
6.1 

Installation overview 

 

 

 
 
6.2 

Converter 

To fix the converter, 2 drill holes for screws 
with 4 mm diameter are to reach after remov-
ing the cover. 

 

Note: Only operate the device for a 
prolonged period of time when it is 
closed! After setting the parameters, 

replace the housing cover immediately to 
prevent dirt from entering the device and to 
meet the electromagnetic requirements (EMV).

 

Cover

Drill holes

Содержание Impac INFRATHERM ISR 50-LO

Страница 1: ...1 ISR 50 LO INFRATHERM Pyrometer Betriebsanleitung User Manual IMPAC Spezialist f r ber hrungslose Temperaturmessung IMPAC Specialist in non contact thermometry www imaxsun com...

Страница 2: ...5 2 1 Short description 36 5 3 Connecting the pyrometer to a PC 36 5 4 Connecting to RS232 interface 37 5 5 Connecting to RS485 interface 37 5 6 Connection of additional analyzing devices 37 6 Mechan...

Страница 3: ...mum internal temperature MaxIntTemp 45 10 Settings via interface and software 45 10 1 Installation 45 10 2 Program start 45 10 3 The menu 46 10 4 Beginning 46 10 5 Number of devices 46 10 6 Basic sett...

Страница 4: ...arameter settings Parameters Adjustable or readable at the instrument or via interface Measuring temperature operation mode ratio mono emissivity slope K or emissivity exposure time t90 clear times tc...

Страница 5: ...late the temperature This technique offers the following advantages compared with the standard one color pyrometers The temperature measurement is independent of the emissivity of the object in wide r...

Страница 6: ...Only qualified specialists are allowed to connect such devices to the mains voltage 5 Electrical Installation The ISR 50 LO is powered by a voltage of 24 V DC possible range 18 36 V DC With the conne...

Страница 7: ...g is frozen as long as pin 7 24 V is connected to pin 5 Pin 8 6 If the pyrometer is used in ratio mode a relay contact for the integrated contamination alarm is available It switches at the too low si...

Страница 8: ...rate has to be reduced by 50 when the transmission distance is doubled see also 9 9 Baud rate 5 6 Connection of additional analyzing devices Additional analyzing instruments for example a LED digital...

Страница 9: ...radius 0 6 mm fibre 0 2 mm fibre 0 1 mm fibre green mark red mark yellow mark for short time max 250 C 30 mm 50 mm 50 mm permanent max 250 C 50 mm 120 mm 120 mm wound up Zustand max 50 C 50 mm 120 mm...

Страница 10: ...ce possible range is 340 4500 mm the measuring distance is always measured from the front of the lens Only in this distance the spot sizes mentioned in the table 7 1 will be achieved A tape can be use...

Страница 11: ...e user of an own communication software find all inter face commands on section 14 Data format UPP 8 1 Settings at the instrument The LC display as well as the push buttons for displaying and setting...

Страница 12: ...Adress Adr 00 Baud rate kBaud 19 2 kBd Wait time tw for RS485 10 Temperature display C F C Switch for interface RS485 RS232 RS232 9 Parameters settings 9 1 Operating mode Mode The standard use for the...

Страница 13: ...same temperature Different materials have different emissivities ranging between 0 and 100 settings at the pyrometer between 5 and 100 Additionally the emissivity is depending on the surface condition...

Страница 14: ...Here the maximum value for each object has to be indicated In this mode the maximum value is stored until a new hot object appears in the measuring beam The temperature which has to be recognized as h...

Страница 15: ...nal in such a high level that a correct temperature measurement will be impossible this can also happen if the object is smaller than the spot size of the pyrometer To avoid wrong measurements in adva...

Страница 16: ...he operating and analyzing software InfraWin is included in delivery of the pyrometer With this software all pyrometer functions also can be used on the PC except changing the interface or using the t...

Страница 17: ...sing RS485 may be operated simultaneously by the same interface if two different addresses have been properly entered see 9 8 Address 10 5 Number of devices With a click on Number of devices InfraWin...

Страница 18: ...ngth of the relative signal warning level graphically and numerically switch off level graphically and numerically The measuring mode the emissivity or emissivity slope and values for the warning leve...

Страница 19: ...nce the temperature entry has been entered and confirmed with OK InfraWin will then calculate the emissivity slope which occurs with the new temperature This is displayed immediately and can be used f...

Страница 20: ...e mouse Selecting the Trend output initially causes all the saved data to be displayed If the data exceeds an amount that can be represented reasonably you may Zoom in on a partial segment using the m...

Страница 21: ...ter is designed for a storage temperature of 20 60 C with non condensing conditions Other kind of storage can damage or malfunction the pyrometer 12 Maintenance 12 1 Safety Attention during pyrometer...

Страница 22: ...are see 10 6 Basic settings Test commands can be exchanged directly with the pyrometer The data exchange occurs in ASCII format with the following transmission parameters The data format is 8 data bit...

Страница 23: ...AdwXX XX dec 99 in 00 switches off this function is the set value for the switch off limit aw Address AAgaXX XX 00 97 00 97 regular device addresses 99 global address with response 98 global address w...

Страница 24: ...ough for this 2 The pyrometer s response will follow after 3 ms at latest 3 If there is no response there is a parity or syntax error and the inquiry has to be repeated 15 Reference numbers 15 1 Refer...

Отзывы: