Operating Instructions
Smart Weather Sensor
G. Lufft Mess- und Regeltechnik GmbH, Fellbach, Germany
107
19.6.5.3 Buffer Assignment Additional Measurement Commands M3 / C3: Air
Pressure
Device configured for measurement values in metric or US units:
Measurement Value
UMB
Channel
Min
Max
Unit
Buffer ‘0’
Abs. Air Pressure(act)
300
300.0
1200.0
hPa
Abs. Air Pressure (min)
320
300.0
1200.0
hPa
Abs. Air Pressure (max)
340
300.0
1200.0
hPa
Abs. Air Pressure (avg)
360
300.0
1200.0
hPa
Air Density (act)
310
0.000
3.000
kg/m3
B
uffer ‘1’
Rel. Air Pressure (act)
305
300.0
1200.0
hPa
Rel. Air Pressure (min)
325
300.0
1200.0
hPa
Rel. Air Pressure (max)
345
300.0
1200.0
hPa
Rel. Air Pressure (avg)
365
300.0
1200.0
hPa
Example: Request with M command
0M3!
00009<CR><LF>
0D0!
0+1001.0+1000.0+1002.0+1001.0+1.119<CR><LF>
0D1!
0+1017.0+1016.0+1018.0+1017.0<CR><LF>
Example: Request with C command
0C3!
000009<CR><LF>
0D0!
0+1001.0+1000.0+1002.0+1001.0+1.119<CR><LF>
0D1!
0+1017.0+1016.0+1018.0+1017.0<CR><LF>