Model 70 System Manual
Ludlum Measurements Inc.
16
May 2021
5.0 SPECIFICATIONS
DETECTORS
Model 70/1 or 70/2 CZT detector:
Volume 0.5 cm
3
(typically 5 x 10 x 10 mm CZT)
Model 70/3 or 70/4 CZT detector:
Volume 1.6 cm
3
(typically 16 x 10 x 10 mm CZT)
Neutron detector (included in Models 70/2 and 70/4):
15 mm x 40 mm, Li6-based
CZT Energy Resolution:
Usually used
137
Cs, the FWHM at 662 keV: 1.8% to 2.5%
ANALOG TO DIGITAL CONVERTER
Type:
Successive Approximation ADC
Number of channels:
1024
Lower Threshold:
Internally adjustable – set at 20 keV
Upper Threshold:
Fixed to Channel 1023, all pulses above this level are cosmic and are
accumulated and displayed in Channel 1024
ADC Offset:
Internally adjustable
Nuclide ID Over-Range Dose Rate:
0.5 mSv/hr
DOSIMETER
Energy Range:
30 keV to 3.0 MeV set by the spectrometer
Measuring Units:
Sv, rem, cps
Gamma Dose Rate:
within 30% accuracy per ANSI N42-48, from 0.05 µSv/h to 3 mSv/h (5 µR/hr
to 300 mR/hr)
Output:
Memory or PC
CLOCK/CALENDAR
Type:
Built-in 24-hour clock, 4-year calendar (including leap year)
DATA STORAGE
Type:
1 GB flash memory, extended memory 16 MB
Capacity:
100,000 samples, including time/date (
Dose
mode) or 500 spectra (
Identify
mode - 1024 channels), including time/date
Stored Data Retrieval:
Via USB and/or Wi-Fi to a computer
DATA OUTPUT
Type:
USB port, Wi-Fi
Содержание 70
Страница 1: ...LUDLUM MODEL 70 SPECTROMETRIC PERSONAL RADIATION DETECTOR IDENTIFIER May 2021 ...
Страница 2: ...Model 70 Spectrometric Personal Radiation Detector Identifier SYSTEM MANUAL May 2021 ...
Страница 3: ......
Страница 6: ...Spectrometric Personal Radiation Detector SPRD Radioisotope Identifier RIID ...
Страница 7: ......
Страница 29: ...Model 70 System Manual Ludlum Measurements Inc 22 May 2021 Radiation dose rate profile chart ...