Model 2360
Technical Manual
Section 4
Ludlum Measurements, Inc.
Page 4-5
July 2016
LIM
: a multi-turn potentiometer used to set the maximum
HV
limit to
2000 Vdc
HV
: a multi-turn potentiometer used to adjust the high-voltage test
reading (0 to 2 kV scale) to correspond with the actual high-voltage
output. The
HV
switch must be depressed during adjustment.
LB
: a multi-turn potentiometer used to adjust the minimum battery
voltage level (2.2 Vdc), corresponding to the low-battery indication on the
meter dial. The
BAT
switch position must be selected during adjustment.