Models 375-31H Area Monitor with Neutron Detector
Section 6
Ludlum Measurements, Inc.
Page 6-1
January 2021
Safety Considerations
Environmental Conditions for Normal Use
Indoor use only
No maximum altitude
Temperature range of -15 to 50
°
C (5 to 122 °F); may be certified for
operation from -40 °C (-40 °F)
Maximum relative humidity of less than 95% (non-condensing)
Mains supply voltage range of 100-240 Vac ,
50/60Hz single phase (less than 100 mA typical, 1 amp max) to wall-
mounted DC adapter supplying 9-12 Vdc
Maximum transient voltage of 1500 Vac
Installation Category II (Overvoltage Category as defined by IEC 1010-1)
Pollution Degree 2 (as defined by IEC 664)
Cleaning Instructions and Precautions
The Model 375-31H may be cleaned externally with bleach wipes or with a
damp cloth, using water, Lysol or alcohol as a wetting agent. Do not
immerse the instrument in any liquid. Observe the following precautions
when cleaning:
1.
Turn the instrument
OFF
and disconnect the instrument power
cord.
2.
Allow the instrument to sit for one minute before cleaning.
3.
Disconnect the detector cable before cleaning the detector.
Section