Model 3030P
Technical Manual
Section 7
Ludlum Measurements, Inc.
Page 46
February 2021
RXn – Read the Crosstalk
This command reads the beta-to-alpha or alpha-to-beta crosstalk. The value
must be divided by 10 to get the correct value. Returns “nnn.”
1 = Reads beta-to-alpha crosstalk.
2 = Reads alpha-to-beta crosstalk.
RY – Read Recycle Mode Status
This command reads the status of the recycle mode. Returns “OFF” or
“ON.”
Rn – Read the Alarm Set Points
This command reads the specified alarm set point. Returns “nnnnnn.”
1 = Read alpha alarm set point.
2 = Read beta alarm set point.
3 = Read alpha + beta alarm set point.
Xn – Set Crosstalk Mode
This command sets crosstalk ON/OFF or returns crosstalk status, where
“n” is a value of 0, 1, or 2.
0 = Sets crosstalk OFF.
1 = Sets crosstalk ON.
2 = Returns status.
SBnxxxxxx – Set Background Subtract
This command sets the alpha or beta background subtract.
1 = Set alpha background subtract.
2 = Set beta background subtract.
SCmmddyyyy – Set Calibration Date
This command sets the calibration due date.
SDmmddyyyyz – Set Date
This command sets the date on the read-time clock. The “z” is the value of
the year Mod 4.
Содержание 251450
Страница 2: ......
Страница 5: ......
Страница 16: ...Model3030P TechnicalManual Section2 Ludlum Measurements Inc Page 8 February 2021 Select Extract Select Next ...
Страница 58: ...Model3030P TechnicalManual Section7 Ludlum Measurements Inc Page 50 February 2021 ...
Страница 59: ...Model3030P TechnicalManual Section7 Ludlum Measurements Inc Page 51 February 2021 ...
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......