Model 3030P
Technical Manual
Section 7
Ludlum Measurements, Inc.
Page 40
February 2021
Clear Samples
: This button clears the instrument of all samples in memory.
Save Samples
: Allows saving the sample data to a user-defined file. The
data is saved in a comma delimited (*.csv) file to allow for importing into
other applications.
Print Samples
: Prints the sample data to a printer. If connected to a PC the
sample data will be immediately printed to the default printer.
Spectral Information
The spectral information tab is useful for generating an alpha spectrum for
an unknown sample. This process is slow; essentially the unit adjusts the
threshold and window and takes a reading at each step within the specified
range. The user can choose the beginning threshold, the region width, and
the ending window set points (all in mV). The user also chooses the count
time for each step. When the count time is chosen, the time switch on the
instrument needs to be set to “PC.” If the 3030P is not set to PC control,
the spectral data acquisition time is determined by the setting on the 3030P.
Pressing the “Read” button in the Spectral Information Tab will confirm the
count time that will be used. A table is generated as the data is collected.
Содержание 251450
Страница 2: ......
Страница 5: ......
Страница 16: ...Model3030P TechnicalManual Section2 Ludlum Measurements Inc Page 8 February 2021 Select Extract Select Next ...
Страница 58: ...Model3030P TechnicalManual Section7 Ludlum Measurements Inc Page 50 February 2021 ...
Страница 59: ...Model3030P TechnicalManual Section7 Ludlum Measurements Inc Page 51 February 2021 ...
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......