![Ludlum Measurements 25009185 Скачать руководство пользователя страница 83](http://html.mh-extra.com/html/ludlum-measurements/25009185/25009185_manual_1946954083.webp)
A.4. SETTINGS
77
CPM:
Mode that automatically transmits the CPM (counts per minute) for the external de-
tector no matter what detector or mode is currently being displayed. The count interval is
based on the AuxCom Auto Mode Interval variable.
Display:
Mode that automatically transmits the numeric value as well as units prefix, units,
and units suffix for the currently displayed screen. The count interval is based on the Aux-
Com Auto Mode Interval variable.
A.4.2 AuxCom Setting - Off Time
This setting defines the number of minutes the AuxCom port will sit idle before turning off.
The primary use for this parameter is to save power while using Bluetooth® by turning off
the Bluetooth® module if there is no connection. The table below shows valid settings.
Setting
Description
0
Disabled
1-9
Inactivity minutes before auto off
Table A.2: AuxCom Auto Off Time
A.4.3 AuxCom Setting - Pwr Mode
This setting determines whether or not the AuxCom module is turned on when the instru-
ment is powered on. In addition to power savings, this setting is also intended for use when
a user has a need for “radio silence” (no RF broadcasts). This setting does not override the
hot key function. Again, the primary use for this parameter is while using Bluetooth®. The
table below shows valid settings.
Description
From Hotkey
At Boot
Table A.3: AuxCom Power on at Boot-up
A.4.4 AuxCom Setting - Write Protect
This setting is an additional security measure that prevents commands that would mod-
ify parameters from being executed over the AuxCom port. While using the instrument
equipped with a wireless radio, such as the Bluetooth® module, there is the potential for
the instrument becoming mis-configured, inadvertently or intentionally, from a remote lo-
cation. This setting is meant to ensure that cannot happen. The table below shows valid
settings.
Setting
Description
0
AuxCom Write Protect Disabled
1
AuxCom Write Protect Enabled
Table A.4: AuxCom Write Protect
,
If write protect is enabled, it can only be disabled by physically connecting it to
the instrument through the USB port.
December 2021
Содержание 25009185
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 12: ......
Страница 14: ......
Страница 32: ......
Страница 35: ...29 Weight 0 7 kg 1 6 lb excluding detectors and clips December 2021 ...
Страница 36: ......
Страница 56: ......
Страница 62: ......
Страница 72: ......
Страница 74: ......
Страница 78: ......