2.8. MAX MODE OPERATION
21
ð
When switching detectors, it is recommended that the current detector be dis-
connected from the instrument to prevent an unintended overvoltage from
damaging the current detector.
If other operational modes are available, a short press of the MODE button will move to the
next available operational mode.
By default, a short press of the LOG button, if datalogging is enabled, will log the currently
displayed rate.
Figure 2.8: RATE Mode Displaying Background Radiation Rate
2.8 MAX Mode Operation
While in MAX mode, the highest detected count rate (since MAX mode was selected) is dis-
played. The word MAX will be displayed when in MAX mode. A short press of the UNITS
button will switch the displayed value between the Primary and Secondary Units.
A short press of the ON/ACK button will cycle through the “click” audio volume levels. If dual
level audio is disabled, a short press of the ON/ACK button will toggle audio between Off and
On. If dual level audio is enabled, a short press of the ON/ACK button will toggle the audio
between "Hi, LO, and OFF". When Sigma Audio is enabled, a short press of the ON/ACK will
toggle the Sigma Audio volume levels.
Under a non-alarm or alarm condition, a long press of the ON/ACK button will reset the
display. When Sigma Audio is enabled, the Sigma Audio alarm level cannot be reset in Max
Mode.
If an alarm condition is present, a short press of the ON/ACK button will acknowledge and
turn off the alarm audio. Under an alarm condition, the ALARM display indicator will be
activated. There are two user settable alarms for both the primary and secondary units. If
both alarms for a particular unit are set, the first alarm will activate beeping alarm audio and
a flashing ALARM display indicator; If only one alarm is set or the second alarm is triggered,
both the alarm audio and the ALARM display indicator are continuous. Alarms are latching
in MAX Mode. Under an alarm condition, the ALARM display indicator will remain on until
the display is reset.
If other detectors are available, a short press of the DETECTOR will switch between available
detectors as evidenced by the DET LEDs.
December 2021
Содержание 25009185
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 12: ......
Страница 14: ......
Страница 32: ......
Страница 35: ...29 Weight 0 7 kg 1 6 lb excluding detectors and clips December 2021 ...
Страница 36: ......
Страница 56: ......
Страница 62: ......
Страница 72: ......
Страница 74: ......
Страница 78: ......