Model 177
Technical Manual
Section 5
Ludlum Measurements, Inc.
Page 5-1
October 2018
Safety Considerations
Environmental Conditions for Normal Use
Indoor use only
No maximum altitude
Temperature range of -20 to 50
°
C (-4 to 122 °F)
Maximum relative humidity of less then 95% (non-condensing)
Wall transformer supply voltage range of 95-250 Vac, 50/60 Hz single
phase (less than 100 mA)
Maximum transient voltage of 1500 Vac
Installation Category II (Overvoltage Category as defined by IEC 1010-1)
Pollution Degree 2 (as defined by IEC 664) (Normally only
nonconductive pollution occurs. Temporary conductivity caused by
condensation is to be expected.)
Cleaning Instructions and Precautions
The Model 177 Alarm Ratemeter may be cleaned externally with a damp
cloth, using only water as the wetting agent. Do not immerse the instrument
in any liquid. Observe the following precautions when cleaning:
1.
Turn the instrument
OFF
and disconnect the instrument power
cord.
2.
Allow the instrument to sit for one minute before cleaning.
Section
5
Содержание 177
Страница 1: ...LUDLUM MODEL 177 ALARM RATEMETER October 2018 Serial Number 322945 and Succeeding Serial Numbers ...
Страница 2: ...LUDLUM MODEL 177 ALARM RATEMETER October 2018 Serial Number 322945 and Succeeding Serial Numbers ...
Страница 4: ......
Страница 5: ......
Страница 38: ......
Страница 39: ......
Страница 40: ......
Страница 41: ......
Страница 42: ......
Страница 43: ......
Страница 44: ......
Страница 45: ......
Страница 46: ......
Страница 47: ......