Model 177
Technical Manual
Section 4
Ludlum Measurements, Inc.
Page 4-4
October 2018
Internal Controls (Overhaul Only)
The following controls are located internally, on the main circuit board:
BAT C:
used to adjust charge voltage to 6.825 volts.
BAT T:
used to adjust meter test voltage reading to 5.97 volts at the
BAT
OK
line.
HV T:
used to adjust the high-voltage test reading to correspond with the
actual high-voltage output.