ELECTRIC SPECIFICATIONS
1. AC Voltage
Range
Resolution
Accuracy
(50 Hz ~ 60 Hz)
Overload
Protection
400.0 mV
0.1 mV
(1.0% + 8 dgts)
1000 Vrms
4.000 V
0.001 V
40.00 V
0.01 V
400.0 V
0.1 V
1000 V
400.0 mV
4.000 V
40.00 V
400.0 V
1000 V
1 V
Input Impedance (nominal) : 10 MΩ, < 100pF
Response : Average RMS
2. DC Voltage
Input Impedance (nominal) : 10 MΩ, < 100pF
3. Resistance
Range
Resolution
Accuracy
Overload
Protection
400.0 Ω
0.1 Ω
(1.0% + 5 dgts)
400 Vrms
4.000 kΩ
0.001 KΩ
40.00 kΩ
0.01 KΩ
400.0 kΩ
0.1 KΩ
4.000 MΩ
0.001 MΩ
40.00 MΩ
0.01 MΩ
(1.5%+10dgts)
Input Impedance (nominal) : 10 MΩ, < 100pF
4. Continuity Test
5. Insulation Test
Short-Circuit Test Current (nominal) : 0.5 mA
Auto Discharge : Discharge time <1 sec. for C ≤ 1 uF
Minimum Measurement : 0.1 MΩ
Range
Resolution
Accuracy
Overload
Protection
0.1 mV
(0.8%+10dgts)
1000 Vrms
0.001 V
0.01 V
0.1 V
1 V
Overload Protection Open Circuit Voltage Threshold (Appx.)
400 Vrms
< 0.44 V
< 40 Ω
Output
Voltage
Display
Range
Resolution Test Current Accuracy
100 V
0.01 ~ 20.00 MΩ
0.01 MΩ
0.5mA
@ 100 KΩ
(3 % + 5)
20.0 ~ 100.0 MΩ
0.1 MΩ
250 V
0.01 ~ 20.00 MΩ
0.01 MΩ
0.5mA
@ 250 KΩ
(3 % + 5)
20.0 ~ 200.0 MΩ
0.1 MΩ
500V
0.01 ~ 20.00 MΩ
0.01 MΩ
0.5mA
@ 500 KΩ
(3 % + 5)
20.0 ~ 200.0 MΩ
0.1 MΩ
200 ~ 500 MΩ
1 MΩ
(5 % + 5)
1000 V
0.01 ~ 20.00 MΩ
0.01 MΩ
0.5mA
@ 1 MΩ
(3 % + 5)
20.0 ~ 200.0 MΩ
0.1 MΩ
200 ~ 2000 MΩ
1 MΩ
(5 % + 5)
Sleep Mode
The meter automatically enters “Sleep mode” after 30 minutes
non -use. The meter comes out of the Sleep mode when a button
is pressed.
The Sleep mode is always disabled in the Hold or Autohold mode
and when performing an insulation test.
Maintenance
Turn instrument off and disconnect test leads.
Clean the instrument by using a damp cloth. Do not use abrasive
cleaners or solvents.
GENERAL SPECIFICATIONS
Operating Temperature
Storage Temperature
Temperature Coefficient
Relative Humidity
Altitude
Safety
: 0 °C to 50 °C (32 °F to 122 °F)
at <75% R.H.
: -20 °C to 60 °C (-4 °F to 140 °F)
at <80% R.H.
: nominal 0.15 x (specified accuracy)
/°C@<18°Cor>28°C(<64°For>82°F),
or otherwise specified
: 0% to 95% @ 10 °C to 30 °C
(50 °F to 86 °F)
0% to 75% @ 30 °C to 40 °C
(86 °F to 104 °F)
0% to 40% @ 40 °C to 50 °C
(104 °F to 122 °F)
: Operating – up to 2000m
Storage – 10000m
: Complies with UL61010-1: 3rd Edition,
Overload protection
Certifications
Battery
CAN / CSA-C22.2 No. 61010-1-12 :
3rd Edition, IEC / EN 61010-1 : 2010 ;
Overload protaction CAT IV 600V and
CAT III 1000V
: CAT IV 600V and CAT III 1000V
: ETL & cETL and CE
: 6 x AAA batteries
(NEDA 24A or IEC LR03)
Battery Life
: Meter use – 1000 hours
Insulation Test use – Meter can perform
at least 1000 tests with alkaline
batteries at room temperature.
These standard tests of 1000 V
into 1 MΩ with a duty cycle of 5
seconds on and 25 seconds off.
Pollution Degree
Dimensions
: 2
: 178mm x 89mm x 48mm
Weight
: 425g