LTR MTI 801 Скачать руководство пользователя страница 2

LOCK

TEST

TEST

If the resistance of the device is below 40 Ω, there 
is a continuity beep tone.
If the resistance of the device is more than 40 Ω, 
there is no beep tone.
This is useful for checking wiring connections and 
operation of switches.

CAUTION

Turn off power and discharge all capacitors on 
circuit to be tested before attempting incircuit 
resistance measurements.
Accurate  measurement  is  not  possible  if 
external or residual voltage is present.

INSULATION TEST

CAUTION

Turn off power and discharge all capacitors on circuit to be

tested before attempting Insulation test.

Insert test leads into V and COM input terminals.

Set the rotary selector to the desired test voltage position. 

Connect the lead probes to the circuit under test.
The display shows

until you press

and a valid

insulation resistance reading is obtained.

Press and hold

TEST

button to start the test.

The LCD shows

(high voltage warning)  symbol  along with

the  resistance  in  MΩ  or GΩ .
The

symbol will be displayed on the LCD until the

TEST

button is released. The meter displays

symbol along

with the maximum resistance for the range when the resistance 
is higher than the maximum display range.

NOTE :

When the

button is pressed before the   

TEST

button, the test remains active until the 
button is pressed again.

LOCK

Release  the

button before disconnecting the probes

from the circuit under test. Then, the circuit will automatically be 
discharged through the meter.

RANGE           AUTOHOLD             LOCK

TEST

SEL

100V

250V

mV

500V

1000V

COM

CAT IV

V

600V

INSULATION

CAT III
1000V

SEL

SEL

OPERATING INSTRUCTIONS

AC/DC Voltage

Set the rotary selector to V 

position.

The meter defaults at AC. 
Press

button momentarily 

to toggle between AC and 
DC.

Insert red lead into V terminal 

and  black  lead  into  COM 

terminal.

Connect black probe to 

ground and red probe to the 

side of the circuit closest to 

the power source.

DC milli-Voltage

Set the rotary selector to 

mV

position.

Insert red lead into V terminal 
and black lead into COM 
terminal.

Connect  black  probe  to 
negative  side  of  the  circuit 
and  red  probe  to  positive 
side  of  the  circuit coming
from the power source.

Resistance

Set the rotary selector to Ω
position.    is displayed.

The meter defaults at Ω

Press

button momentarily

to select

.

Insert black lead into COM 
terminal  and  red  lead into Ω
terminal.

Touchtheprobesacrossthe 
resistance  or  circuit  to be
tested.

RANGE           AUTOHOLD             LOCK

TEST

SEL

100V

250V

mV

500V

1000V

COM

CAT IV

V

600V

INSULATION

CAT III
1000V

RANGE          AUTOHOLD            LOCK

TEST

SEL

100V

250V

mV

500V

1000V

COM

CATIV

V

600V

INSULATION

CAT III

1000V

RANGE           AUTOHOLD             LOCK

TEST

SEL

100V

250V

mV

500V

1000V

COM

CAT IV

V

600V

INSULATION

CAT III
1000V