16 OE15 Membrane Replacement
79
16 OE15 Membrane Replacement
The electrode membrane is very delicate and may be damaged by careless use.
Inspect the electrode periodically for signs of membrane damage and replace it if
necessary.
Remove the damaged membrane and
the two ’O’ rings retaining it and set
these aside. Cut a 120mm length of
membrane and fit to the applicator as
shown by passing it through the middle
and evenly stretching the front edge
back over the outside. Slide the
applicator carefully over the electrode
taking care not to stretch the
membrane. Release the membrane at
the top and withdraw the applicator.
Содержание MOD53
Страница 1: ...MOD53 DISSOLVED OXYGEN MONITOR OPERATION GUIDE...
Страница 26: ...4 Installation Surface Mount 25 Figure 24 ProcessProbeTM Cable...
Страница 37: ...5 User Interface 36 This page intentionally blank...
Страница 41: ...6 Main Display 40 This page intentionally blank...
Страница 47: ...7 Parameters 46 This page intentionally blank...
Страница 57: ...9 Current Output 56 This page intentionally blank...
Страница 71: ...12 Pressure Input 70 This page intentionally blank...
Страница 79: ...15 Fault Finding 78 This page intentionally blank...
Страница 83: ...17 Guarantee and Service 82 This page intentionally blank...
Страница 87: ...Appendix B Probe Parameters 86 This page intentionally blank...
Страница 91: ...Appendix C Factory Default Setups 90 This page intentionally blank...
Страница 95: ...Appendix D Customer Setup 94 This page intentionally blank...
Страница 99: ...Appendix F Temperature Sensor Data 98 This page intentionally blank...
Страница 101: ...Appendix G Pressure Conversions 100 This page intentionally blank...
Страница 107: ...Appendix I Error Messages 106 This page intentionally blank...