10 Calibration
61
TEMPERATURE CALIBRATION
To initiate the temperature calibration select the “
Temperature
“ menu item in the
“
Calibration
” menu and press the “ENTER” key.
Temperature
Temperature offset calibration
(Only present if Compensation is Automatic)
Are You Sure?
The instrument will ask for confirmation
Press to confirm, any of the other 3 buttons
abandons the temperature calibration.
25.0
25.0
25.0
25.0
°C
The Temperature reading is now displayed.
Whilst the display flashes
Ø
Ø
Ø
Ø×
×
×
×
Adjust the reading using the up/down buttons
25.2
25.2
25.2
25.2
°C
Adjusted Reading
Press to accept the calibrated value when
ready.
Offset = +0.2°C
For user information, the unit will display the
offset adjustment
Содержание MOD53
Страница 1: ...MOD53 DISSOLVED OXYGEN MONITOR OPERATION GUIDE...
Страница 26: ...4 Installation Surface Mount 25 Figure 24 ProcessProbeTM Cable...
Страница 37: ...5 User Interface 36 This page intentionally blank...
Страница 41: ...6 Main Display 40 This page intentionally blank...
Страница 47: ...7 Parameters 46 This page intentionally blank...
Страница 57: ...9 Current Output 56 This page intentionally blank...
Страница 71: ...12 Pressure Input 70 This page intentionally blank...
Страница 79: ...15 Fault Finding 78 This page intentionally blank...
Страница 83: ...17 Guarantee and Service 82 This page intentionally blank...
Страница 87: ...Appendix B Probe Parameters 86 This page intentionally blank...
Страница 91: ...Appendix C Factory Default Setups 90 This page intentionally blank...
Страница 95: ...Appendix D Customer Setup 94 This page intentionally blank...
Страница 99: ...Appendix F Temperature Sensor Data 98 This page intentionally blank...
Страница 101: ...Appendix G Pressure Conversions 100 This page intentionally blank...
Страница 107: ...Appendix I Error Messages 106 This page intentionally blank...