Faults
- 66 -
MXD70 pH / Redox
Setup and Operating Guide
Faults
The Sensor Reading Is Incorrect
If reading pH, check that the correct probe type has been selected (see pH / Redox Input Channel
Setup, page 23).
Check that no error messages are being displayed. Check that the sensor cable has been correctly
connected (see Installation Section, Page 7).
Check that the Temperature reading is correct.
Check the instrument calibration using a pH simulator, Adjust the channel calibration if necessary
(see Calibration Section).
Use another instrument to check the sensor.
The Sensor Is Not Functioning Correctly
Check that the sensor glass is not broken or cracked.
Check the reference probe KCl (where applicable) for leakage or contamination.
Ensure all probe protective caps have been removed.
Check that any junction boxes used are correctly connected.
Check that a suitable high impedance, low noise cable has been used.
Check for damaged or broken cables.
Check for damp, grease, or liquids near connectors, junction boxes, or terminations.
Where extension cables have been used, try connecting the sensor directly to the instrument.
The Temperature Reading Is Incorrect
Check that the temperature sensor is correctly attached. (Installation Section, page 16).
Check that the temperature sensor type is correctly selected in the Channel Setup menu (See page
23)
Where practical check the temperature sensor resistance against the table on page 46.
Current Output Is Incorrect or Noisy
Check that the maximum load for the current loop has not been exceeded. (750
).
Check that the terminals have been wired correctly.
Check that the cable screen is attached to Earth at one end and that the cable does not pass too
close to a power cable.
Check that he current output has been configured properly.
Relays Appear to Malfunction
Check that the unit is “On-Line” (see page 23)
Check that the set point configuration is correct (see Setpoints, Current Outputs and Digital Input
Configuration Guide)
If the relays are vibrating or “chattering” as they pass the set point, check the hysteresis setting and
increase if necessary.
Ensure that the relays are connected properly (see MXD73 Installation guide or MXD75 Installation
guide) and that the voltage/current levels are not exceeding
5A @ 30V DC or 5A @ 250V AC.
Check that the instrument input cables are not picking up excessive noise.
Problems with Cables and Connectors
The cable connecting the pH probe to the instrument is the most common cause of problems in pH
measurement systems. The importance of the following recommendations cannot be over stressed.
Input Resistance
The high input resistance, required for the optimum performance from a pH electrode system, will be
seriously degraded if any grease, dirt, or dampness is allowed to collect around any of the connections
between the probe and the instrument. This includes any connectors or junction boxes which may be
used. Particular attention must be paid to the method of extending the pH electrode cable. A general
rule would be the fewer connections there are, the less problems are likely to occur.
Содержание MXD70 Series
Страница 6: ...MXD73 Multi parameter Monitor Installation Guide ...
Страница 7: ......
Страница 10: ...Preface MXD73 Installation Guide 3 Declaration of Conformity ...
Страница 11: ...Preface 4 MXD73 Installation Guide Blank ...
Страница 13: ...Contents 6 MXD73 Installation Guide Blank ...
Страница 15: ...Introduction 8 MXD73 Installation Guide Introduction MXD73 Rear Termination Overview ...
Страница 25: ...Installation 18 MXD73 Installation Guide Installation Blank ...
Страница 33: ...Notes 26 MXD73 Installation Guide ...
Страница 34: ......
Страница 36: ...MXD75 Multi parameter Monitor Installation Guide ...
Страница 37: ......
Страница 40: ...Preface MXD75 Installation Guide 3 Declaration of Conformity ...
Страница 41: ...Preface 4 MXD75 Installation Guide Blank ...
Страница 43: ...Contents 6 MXD75 Installation Guide Blank ...
Страница 45: ...Introduction 8 MXD75 Installation Guide Introduction MXD75 Termination Overview ...
Страница 58: ...Installation MXD75 Installation Guide 21 Installation Horizontal Mounting Vertical Mounting ...
Страница 59: ...Installation 22 MXD75 Installation Guide Installation Blank ...
Страница 67: ...Notes 30 MXD75 Installation Guide ...
Страница 68: ...Notes MXD75 Installation Guide 31 ...
Страница 70: ......
Страница 72: ...MXD70 SERIES Multi parameter Monitor User Interface Data Logging Guide ...
Страница 73: ......
Страница 77: ...Contents 4 User Interface Guide Blank Page ...
Страница 81: ...User Interface 8 User Interface Guide User Interface Blank Page ...
Страница 85: ...Security Code Access 12 User Interface Guide Security Code Access Blank Page ...
Страница 93: ...Update Software 20 User Interface Guide Update Software Blank Page ...
Страница 107: ...Data Logging 34 User Interface Guide Data Logging Blank Page ...
Страница 109: ...Error Messages 36 User Interface Guide Error Messages Blank Page ...
Страница 116: ......
Страница 117: ......
Страница 118: ...MXD70 SERIES Multi parameter Monitor Conductivity Setup and Operating Guide ...
Страница 119: ......
Страница 123: ...Contents 4 MXD70 Conductivity Setup and Operating Guide Blank ...
Страница 133: ...Installation 14 MXD70 Conductivity Setup and Operating Guide Installation Blank ...
Страница 149: ...Channel Calibration 30 MXD70 Conductivity Setup and Operating Guide Channel Calibration Blank ...
Страница 155: ...Appendix 36 MXD70 Conductivity Setup and Operating Guide Appendix ...
Страница 177: ...Notes 58 MXD70 Conductivity Setup and Operating Guide ...
Страница 178: ...Notes MXD70 Conductivity Setup and Operating Guide 59 ...
Страница 180: ......
Страница 181: ......
Страница 182: ...MXD70 SERIES Multi parameter Monitor pH Redox Setup and Operating Guide ...
Страница 183: ......
Страница 211: ...Channel Setup 28 MXD70 pH Redox Setup and Operating Guide Channel Setup Blank ...
Страница 251: ...Notes 68 MXD70 pH Redox Setup and Operating Guide ...
Страница 252: ...Notes MXD70 pH Redox Setup and Operating Guide 69 ...
Страница 254: ......
Страница 256: ...MXD70 SERIES Multi parameter Monitor Electrodeless Conductivity Setup and Operating Guide ...
Страница 257: ......
Страница 261: ...Contents 4 MXD70 Electrodeless Conductivity Setup and Operating Guide Blank ...
Страница 263: ...Specification 6 MXD70 Electrodeless Conductivity Setup and Operating Guide Specification Blank ...
Страница 295: ......
Страница 311: ...Notes 54 MXD70 Electrodeless Conductivity Setup and Operating Guide ...
Страница 313: ......
Страница 314: ......
Страница 315: ...MXD70 SERIES Multi parameter Monitor Dissolved Oxygen Setup and Operating Guide ...
Страница 316: ......
Страница 320: ...Contents 4 MXD70 Dissolved Oxygen Setup and Operating Guide Blank ...
Страница 322: ...Specification 6 MXD70 Dissolved Oxygen Setup and Operating Guide Specification Blank ...
Страница 339: ...Channel Setup MXD70 Dissolved Oxygen Setup and Operating Guide 23 Channel Setup Blank ...
Страница 353: ...Channel Calibration MXD70 Dissolved Oxygen Setup and Operating Guide 37 Channel Calibration ...
Страница 359: ...Appendix MXD70 Dissolved Oxygen Setup and Operating Guide 43 Appendix Blank ...
Страница 381: ...Notes MXD70 Dissolved Oxygen Setup and Operating Guide 65 ...
Страница 382: ...Notes 66 MXD70 Dissolved Oxygen Setup and Operating Guide ...
Страница 384: ......
Страница 385: ......
Страница 386: ...MXD70 SERIES Multi parameter Monitor Auxiliary mA Input Setup and Operating Guide ...
Страница 387: ......
Страница 391: ...Contents 4 MXD70 Auxiliary mA Input Setup and Operating Guide Blank ...
Страница 393: ...Specification 6 MXD70 Auxiliary mA Input Setup and Operating Guide Specification Blank ...
Страница 405: ...Channel Setup 18 MXD70 Auxiliary mA Input Setup and Operating Guide Channel Setup BLANK ...
Страница 413: ...Channel Calibration 26 MXD70 Auxiliary mA Input Setup and Operating Guide Channel Calibration Blank ...
Страница 434: ...Notes MXD70 Auxiliary mA Input Setup and Operating Guide 47 ...
Страница 436: ......
Страница 437: ......
Страница 438: ...MXD70 SERIES Multi parameter Monitor Suspended Solids Turbidity Setup and Operating Guide ...
Страница 439: ......
Страница 443: ...Contents 4 MXD70 Suspended Solids Turbidity Setup and Operating Guide Blank ...
Страница 445: ...Specification 6 MXD70 Suspended Solids Turbidity Setup and Operating Guide Specification Blank ...
Страница 463: ...Channel Calibration 24 MXD70 Suspended Solids Turbidity Setup and Operating Guide Channel Calibration Blank ...
Страница 465: ...Appendix 26 MXD70 Suspended Solids Turbidity Setup and Operating Guide Appendix Blank ...
Страница 486: ...Notes MXD70 Suspended Solids Turbidity Setup and Operating Guide 47 ...
Страница 488: ......
Страница 489: ......
Страница 490: ...MXD70 SERIES Multi parameter Monitor Setpoints Current Outputs And Digital Inputs Configuration Guide ...
Страница 491: ......
Страница 495: ...Contents 4 MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide Blank ...
Страница 511: ...Setpoints 20 MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide Setpoints ...
Страница 519: ...Current Outputs 28 MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide Current Outputs Blank ...
Страница 525: ...Notes 34 MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide ...
Страница 526: ...Notes MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide 35 ...
Страница 528: ......
Страница 529: ......
Страница 530: ...MXD70 Series Multi parameter Monitor Modbus RS485 Interface Operating Guide ...
Страница 531: ......
Страница 535: ...Contents 4 Modbus RS485 Interface Operating Guide BLANK ...
Страница 539: ...Modbus RS485 8 Modbus RS485 Interface Operating Guide Modbus RS485 BLANK ...
Страница 628: ...Notes Modbus RS485 Interface Operating Guide 97 ...
Страница 629: ...Notes 98 Modbus RS485 Interface Operating Guide ...
Страница 630: ......