Channel Calibration
MXD70 pH / Redox
Setup and Operating Guide
- 29 -
Channel Cal
ibration
Calibration
Calibration Procedures
Normal good practices should be observed when calibrating electrode systems against standard
solutions.
Always clean the electrode system before inserting in the standard solution to avoid
contamination and to obtain the best response from the electrode system.
Rinse the electrode system in clean (preferably demineralised) water between standard solutions.
Allow time for the electrode system temperature to stabilise in each standard solution.
Use standard solutions of known quality. If in doubt make up fresh solutions.
Use clean beakers and bottles for standard solutions.
The input channel can be taken offline, which de-energises the relays and holds the current
outputs at their last values. This facility is useful when calibrating the system, however the
operator must ensure the relays are in a safe state when using this feature.
pH Buffers
LTH supplies buffer powders which are prepared from pure dry chemicals, sealed in a sachet or capsule
which prevents the absorption of moisture and subsequent deterioration. Buffer solutions should be
prepared in accordance with the instructions supplied with them.
The MXD70 Series offers two different but simple methods of calibration for pH electrode systems.
Auto
In Auto calibration the instrument requires that the electrode is first placed in a 4pH buffer and then
secondly in a 9pH buffer. These are nominal values for which the instrument carries the exact LTH
buffer values due to temperature variations (other pH buffer solution values at temperature can also
be entered). From these two points the instrument then calculates the Offset and Slope for the
electrode. NB. Auto calibration mode is not available when the pH probe type is set to antimony.
In order for the auto calibration to work correctly, the buffer temperature must either be measured by
the instrument during calibration, or if manual temperature compensation is being used the buffer
temperature must be entered in the Calibration Manual Temperature Input in the calibration menu.
Manual
In Manual calibration mode it is possible to do single or two point calibration, using either the
combined
Buffer
and
Slope
menu or individual Buffer menu. It is important to do the calibration at
suitable values i.e.
Adjustment of the Buffer (Offset) value at 7.00pH
Adjustment of the Slope (Gain) value at any other pH (usually 4 or 9 pH).
The output from an ideal glass/reference electrode pair will normally be 0mV at 7pH, therefore the
slope will have little or no effect at 7pH but increasing influence the further from 7pH the calibration
point is.
If a single point calibration is required the buffer should be adjusted. As actual pH buffer values are
used no compensation is made for the buffer solution with temperature in the instrument. It is
important therefore to note the actual buffer value at the temperature of the solution, see page 33 for
a table showing pH variations versus temperature for LTH’s standard solutions.
Содержание MXD70 Series
Страница 6: ...MXD73 Multi parameter Monitor Installation Guide ...
Страница 7: ......
Страница 10: ...Preface MXD73 Installation Guide 3 Declaration of Conformity ...
Страница 11: ...Preface 4 MXD73 Installation Guide Blank ...
Страница 13: ...Contents 6 MXD73 Installation Guide Blank ...
Страница 15: ...Introduction 8 MXD73 Installation Guide Introduction MXD73 Rear Termination Overview ...
Страница 25: ...Installation 18 MXD73 Installation Guide Installation Blank ...
Страница 33: ...Notes 26 MXD73 Installation Guide ...
Страница 34: ......
Страница 36: ...MXD75 Multi parameter Monitor Installation Guide ...
Страница 37: ......
Страница 40: ...Preface MXD75 Installation Guide 3 Declaration of Conformity ...
Страница 41: ...Preface 4 MXD75 Installation Guide Blank ...
Страница 43: ...Contents 6 MXD75 Installation Guide Blank ...
Страница 45: ...Introduction 8 MXD75 Installation Guide Introduction MXD75 Termination Overview ...
Страница 58: ...Installation MXD75 Installation Guide 21 Installation Horizontal Mounting Vertical Mounting ...
Страница 59: ...Installation 22 MXD75 Installation Guide Installation Blank ...
Страница 67: ...Notes 30 MXD75 Installation Guide ...
Страница 68: ...Notes MXD75 Installation Guide 31 ...
Страница 70: ......
Страница 72: ...MXD70 SERIES Multi parameter Monitor User Interface Data Logging Guide ...
Страница 73: ......
Страница 77: ...Contents 4 User Interface Guide Blank Page ...
Страница 81: ...User Interface 8 User Interface Guide User Interface Blank Page ...
Страница 85: ...Security Code Access 12 User Interface Guide Security Code Access Blank Page ...
Страница 93: ...Update Software 20 User Interface Guide Update Software Blank Page ...
Страница 107: ...Data Logging 34 User Interface Guide Data Logging Blank Page ...
Страница 109: ...Error Messages 36 User Interface Guide Error Messages Blank Page ...
Страница 116: ......
Страница 117: ......
Страница 118: ...MXD70 SERIES Multi parameter Monitor Conductivity Setup and Operating Guide ...
Страница 119: ......
Страница 123: ...Contents 4 MXD70 Conductivity Setup and Operating Guide Blank ...
Страница 133: ...Installation 14 MXD70 Conductivity Setup and Operating Guide Installation Blank ...
Страница 149: ...Channel Calibration 30 MXD70 Conductivity Setup and Operating Guide Channel Calibration Blank ...
Страница 155: ...Appendix 36 MXD70 Conductivity Setup and Operating Guide Appendix ...
Страница 177: ...Notes 58 MXD70 Conductivity Setup and Operating Guide ...
Страница 178: ...Notes MXD70 Conductivity Setup and Operating Guide 59 ...
Страница 180: ......
Страница 181: ......
Страница 182: ...MXD70 SERIES Multi parameter Monitor pH Redox Setup and Operating Guide ...
Страница 183: ......
Страница 211: ...Channel Setup 28 MXD70 pH Redox Setup and Operating Guide Channel Setup Blank ...
Страница 251: ...Notes 68 MXD70 pH Redox Setup and Operating Guide ...
Страница 252: ...Notes MXD70 pH Redox Setup and Operating Guide 69 ...
Страница 254: ......
Страница 256: ...MXD70 SERIES Multi parameter Monitor Electrodeless Conductivity Setup and Operating Guide ...
Страница 257: ......
Страница 261: ...Contents 4 MXD70 Electrodeless Conductivity Setup and Operating Guide Blank ...
Страница 263: ...Specification 6 MXD70 Electrodeless Conductivity Setup and Operating Guide Specification Blank ...
Страница 295: ......
Страница 311: ...Notes 54 MXD70 Electrodeless Conductivity Setup and Operating Guide ...
Страница 313: ......
Страница 314: ......
Страница 315: ...MXD70 SERIES Multi parameter Monitor Dissolved Oxygen Setup and Operating Guide ...
Страница 316: ......
Страница 320: ...Contents 4 MXD70 Dissolved Oxygen Setup and Operating Guide Blank ...
Страница 322: ...Specification 6 MXD70 Dissolved Oxygen Setup and Operating Guide Specification Blank ...
Страница 339: ...Channel Setup MXD70 Dissolved Oxygen Setup and Operating Guide 23 Channel Setup Blank ...
Страница 353: ...Channel Calibration MXD70 Dissolved Oxygen Setup and Operating Guide 37 Channel Calibration ...
Страница 359: ...Appendix MXD70 Dissolved Oxygen Setup and Operating Guide 43 Appendix Blank ...
Страница 381: ...Notes MXD70 Dissolved Oxygen Setup and Operating Guide 65 ...
Страница 382: ...Notes 66 MXD70 Dissolved Oxygen Setup and Operating Guide ...
Страница 384: ......
Страница 385: ......
Страница 386: ...MXD70 SERIES Multi parameter Monitor Auxiliary mA Input Setup and Operating Guide ...
Страница 387: ......
Страница 391: ...Contents 4 MXD70 Auxiliary mA Input Setup and Operating Guide Blank ...
Страница 393: ...Specification 6 MXD70 Auxiliary mA Input Setup and Operating Guide Specification Blank ...
Страница 405: ...Channel Setup 18 MXD70 Auxiliary mA Input Setup and Operating Guide Channel Setup BLANK ...
Страница 413: ...Channel Calibration 26 MXD70 Auxiliary mA Input Setup and Operating Guide Channel Calibration Blank ...
Страница 434: ...Notes MXD70 Auxiliary mA Input Setup and Operating Guide 47 ...
Страница 436: ......
Страница 437: ......
Страница 438: ...MXD70 SERIES Multi parameter Monitor Suspended Solids Turbidity Setup and Operating Guide ...
Страница 439: ......
Страница 443: ...Contents 4 MXD70 Suspended Solids Turbidity Setup and Operating Guide Blank ...
Страница 445: ...Specification 6 MXD70 Suspended Solids Turbidity Setup and Operating Guide Specification Blank ...
Страница 463: ...Channel Calibration 24 MXD70 Suspended Solids Turbidity Setup and Operating Guide Channel Calibration Blank ...
Страница 465: ...Appendix 26 MXD70 Suspended Solids Turbidity Setup and Operating Guide Appendix Blank ...
Страница 486: ...Notes MXD70 Suspended Solids Turbidity Setup and Operating Guide 47 ...
Страница 488: ......
Страница 489: ......
Страница 490: ...MXD70 SERIES Multi parameter Monitor Setpoints Current Outputs And Digital Inputs Configuration Guide ...
Страница 491: ......
Страница 495: ...Contents 4 MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide Blank ...
Страница 511: ...Setpoints 20 MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide Setpoints ...
Страница 519: ...Current Outputs 28 MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide Current Outputs Blank ...
Страница 525: ...Notes 34 MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide ...
Страница 526: ...Notes MXD70 Setpoints Current Outputs and Digital Inputs Configuration Guide 35 ...
Страница 528: ......
Страница 529: ......
Страница 530: ...MXD70 Series Multi parameter Monitor Modbus RS485 Interface Operating Guide ...
Страница 531: ......
Страница 535: ...Contents 4 Modbus RS485 Interface Operating Guide BLANK ...
Страница 539: ...Modbus RS485 8 Modbus RS485 Interface Operating Guide Modbus RS485 BLANK ...
Страница 628: ...Notes Modbus RS485 Interface Operating Guide 97 ...
Страница 629: ...Notes 98 Modbus RS485 Interface Operating Guide ...
Страница 630: ......