Error message
Status
Possible cause
Remedy
A signal sounds.
Therapy is
interrupted.
Container is not
connected.
Connections open.
The container and exudate pouch may
not be connected to the device properly.
The cause may also be a collapsed
exudate pouch.
Or a connection between the tubes
may be open.
Please check the points above and
re-establish all connections if necessary.
A signal sounds.
Therapy continues
to run in the
background.
Wound dressing
or connections
not tightly sealed.
The tubes might not be connected
properly or the dressing itself is not
sealed tight.
Please check both possibilities,
reconnect the tubes using
the single-piece connector or
improve the seal on the dressing.
A signal sounds.
Therapy continues
to run in the
background.
Exudate flow
interrupted, tube
blocked, internal
filter blocked.
A tube clamp may have accidentally
been closed or there may be a blockage
in the tubes.
Please open all the tube clamps or try
to remove the blockage in the tubes.
If necessary, change the full exudate
pouch.
However, if the internal charcoal/bacterial
filter is blocked, the device needs to
be reprocessed properly by L&R or
an authorised L&R service partner!
A signal sounds.
Therapy continues
to run in the
background.
If the message “Exudate pouch is full”
appears, it is high time to insert a new
exudate pouch. Replace the pouch in
accordance with the instructions for use
(see Section 3.4).
Содержание Suprasorb CNP P3
Страница 2: ......
Страница 4: ......
Страница 5: ...Suprasorb CNP P3 Gebrauchsanweisung ...
Страница 6: ...6 ...
Страница 57: ...Suprasorb CNP P3 Mode d emploi ...
Страница 58: ...58 ...
Страница 109: ...Suprasorb CNP P3 Instructions for use ...
Страница 110: ...110 ...
Страница 161: ...4026372 2018 02 Lohmann Rauscher International GmbH Co KG Westerwaldstraße 4 56579 Rengsdorf Germany 0123 Stand 07 2017 ...