APPENDIX: ELECTROMAGNETIC COMPATIBILITY
Your Cellu M6
®
Endermolab i device requires special care concerning the EMC; it must be
installed and service according to the information provided in this user guide.
Portable and mobile RF communication devices must not be used within 30 cm of your device,
they can cause undesired operation.
The use of other treatment heads than those provided by LPG Systems may result in increased
emissions or decreased immunity of the device.
Your Cellu M6
®
Endermolab i device should not be used adjacent to or stacked with other
medical devices.
The Cellu M6
®
Endermolab i device does not manage essential performances.
Interference may occur near equipment marked with the following symbol:
Conducted RF disturbances
IEC 61000-4-6
3 V 0.15 MHz – 80 MHz
6 V in ISM bands and amateur radio bands between
0.15 MHz and 80 MHz
80% AM to 1 kHz
3 V 0.15 MHz – 80 MHz
6 V in ISM bands and amateur radio bands
between 0.15 MHz and 80 MHz
80% AM to 1 kHz
Power frequency magnetic field
IEC 61000-4-8
30A/m
30 A/m
Voltage Dips and Interruptions:
IEC 61000-4-11
0% UT; 0.5 cycle
A 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0% UT; 1 cycle to 0°
70% UT; 25/30 cycles to 0°
0% UT; 250/300 cycles
0% UT; 0.5 cycle
A 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0% UT; 1 cycle to 0°
70% UT; 25/30 cycles to 0°
0% UT; 250/300 cycles