Sensor LeoCap
|
Rev. 04
Recommended stages for automated cleaning and disinfection
Stage
Temperature
Duration
Pre-rinsing
Cold water
3 min
Cleaning
65 °C
5 min
Neutralizing
Hot water
1 min
Rinsing
Cold distilled water
3 min
Thermal disinfection
93 °C
44 min
Drying
120 °C
12 min
Packaging
After cleaning and disinfection, put the adapter in sterile packaging. The sterile
packaging must fulfil the requirements of the ISO 11607 standard.
Sterilization
Sterilize the adapters in a hot steam sterilizer at 134 °C and 0.21 MPa. Minimum
duration of a sterilization cycle: 5 minutes
6
Storage
Store the device in its original packaging in a closed, ventilated room at an air
temperature between 5 °C and 40 °C and a maximum relative humidity of 80%
at a temperature of 25 °C.
In case of temporary withdrawal from the packaging, store the device at an am-
bient temperature of 5 °C to 40 °C and a maximum relative humidity of 80% (at
25 °C). Devices should be stored on a rack in a single line. Do not store the device
in places where acid vapours or other harmful vapours may occur.
In case of long-term storage after use, the device should be placed in a polyeth-
ylene case and, if possible, in the original packaging to protect it from mechanical
damage.
The device may be transported in the original packing without any limitation of
distance, according to the applicable transportation regulations. The temperature
range for transport is from -50 °C to +50 °C.
Caution
After transport at temperatures below 0 °C, the device should be
allowed to accommodate to the normal ambient conditions for 12
hours before unpacking.
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...