A 5 | Ventilation modes and parameters
102
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
A 5.5.2
Dynamic BiLevel ST
Dynamic BiLevel Spontaneous Timed.
Combination of BiLevel ST mode and dynamic BiLevel with dynamic adjustment
of the mandatory ventilation pressure and the possibility of spontaneous breath-
ing during the entire ventilation cycle.
Breath cycles are defined by setting a fixed respiratory rate (60/f). If spontaneous
breathing results in the machine's trigger threshold being reached, the spontane-
ous breath will be supported with the set pressure. If the patient fails to breathe
spontaneously during the cycle, the machine will initiate a conventional BiLevel
breath with volume guarantee (= dynamic BiLevel) at the end of the calculated
cycle time. In patients with no spontaneous breathing activity, ventilatory support
is delivered in the form of the set mandatory minute volume (f x Vt). In patients
who stop breathing for short periods of time, the mandatory ventilation compo-
nents provide an intelligent back-up system for apnoea ventilation.
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...