12 | Technical specifications
202
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
12.10
Guidance and manufacturer's declaration
Electromagnetic emissions:
Based on the device characteristics determined by its EMISSIONS, the device can be used
in industrial areas and hospitals (CISPR 11 class A). If used in a residential environment (for
which CISPR 11 class B is normally required), this device might not offer adequate protec
-
tion to radio-frequency communication services. The user might need to take mitigation
measures, such as relocating or re-orienting the device.
Emission tests
Compliance
Electromagnetic
environment – guidance
RF emissions
to CISPR 11
Harmonic emissions
to IEC 61000-3-2
Voltage fluctuations/flicker
emissions to IEC 61000-3-3
Class A, group 1
Class A
Complies
The system uses RF energy
only for its internal function.
Therefore, its RF emissions are
very low and are not likely to
cause any interference in near-
by electronic equipment.
Electromagnetic immunity:
elisa 300 / elisa 500 is intended for use in the electromagnetic environment specified be
-
low. It is the responsibility of the customer or user to ensure that elisa 300 / elisa 500 is
operated in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment –
guidance
Electrostatic dis-
charge
(ESD) to
IEC 61000-4-2
± 8 kV
contact
± 15 kV
air
± 8 kV
contact
± 15 kV
air
Floors should be wood,
concrete or ceramic
tile. If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Electrical fast
transients/bursts to
IEC 61000-4-4
± 2 kV for
power supply lines
± 1 kV for input/out-
put lines
± 2 kV for
power supply lines
± 1 kV for input/out-
put lines
The mains power quality
should be that of a
typical commercial or
hospital
environment.
Surges to IEC 61000-
4-5
± 1 kV
differential mode
± 2 kV common mode
± 1 kV
differential mode
± 2 kV common mode
The mains power
quality should be that of
a typical commercial or
hospital environment.
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...