26
TurbiDirect_6 02/2019
11. Press [Read / Avg] or [Read] key.
12. Note the displayed result.
Repeat this procedure until the smallest NTU value is obtained.
13. Mark the vial.
14. To index more vials, repeat points 1. to 13. for each vial.
15. Repeat this procedure until the reading matches the first vial reading within ± 0,01
NTU.
16. Mark the vial.
17. Repeat this procedure to match other vials.
Note:
It may not be possible to match all vials due to variation in the glass.