12
8.6 Input of TDS factor
0.40 ... 1.00:
Conversion factor for TDS measure-
ment.
The conversion factor depends on the composition of
the medium and needs to be determined separately
for each type of water.
8.7 Cell Constant 1/cm
The cell constant may change due to soiling, deposits
or other mechanical influences on the measuring cell.
If an exact standard reference solution is available,
the measuring cell can be adjusted accordingly by
changing the cell constants. Additionally, the cell
constant can be automatically recorded by means of
calibration.
• User calibration:
cELL
• Manufacturing calibration:
CELL
8.8 Selection of temperature
unit °C /°F
°C:
All temperature values in degree Celsius
°F:
All temperature values in degree Fahrenheit
8.9 Zero point temperature offset
-2.0°C...2.0°C bzw. –3.6°F...3.6°F
The zero point for temperature measurement is offset
by this value.
This enables the user to balance out any occurring
sensor
deviations: