LRF-3300S Ultrasonic Transit-time Flow meter
- 13 -
Top View
3.1.4
Z Method
The signal transmitted in a Z method installation has less attenuation than a signal transmitted with the V method
When the pipes are too large, there are some suspended solid in the fluid, or the scaling and liner are too thick.
This is because the Z method utilizes a directly transmitted (rather than reflected) signaling which transverses the
liquid only once. The Z method is able to measure on pipe diameters ranging from 100mm to 5000mm (4 inch to
200 inch) approximately. Therefore, we recommend the Z method for pipe diameters over 300mm (12 inch).
Side view
Section
Top View