Please refer to page 4 for detailed external control function.
Make sure the power line of the pump is grounded reliably to
protect the operator in humid condition.
1
2
Pump Head
Drive
Fuse Socket
External Control Port
Power Socket
External Control Switch
Introduction
Acceptable pump heads for WT600-2J are YZ1515x, YZ2515x, YZII15,
YZII25, BZ25, KZ25, DG15-24 and DMD25. The pump delivers flow rates
from 4.2 to 6000 mL/min. Driven by DC brushless motor the pump has
features of high efficiency, low vibration, high torque and maintenance-
free. It can drive double pump heads. The speed can be adjusted manually
or automatically through external control. The external control
includes analog control and communication control.
Basic Operation
Speed Adjusting Knob
Prime Key
CCW/STOP/CW Switch
Power Switch
LED Display
Speed Adjusting Knob
In manual control mode, turn the
Speed Adjusting Knob
clockwise the speed increase; Turn the
Speed Adjusting Knob
counter clockwise the speed decrease.Turn the
Speed Adjusting
Knob
rapidly the increment or decrement of the speed is 10 rpm.
The pump speed is from 60 rpm to 600 rpm.
Prime Key
In manual control mode press down the
Prime Key
to enter prime
status when the pump runs at full speed (that is 600 rpm) and the
LED displays the max. speed for empting, filling and rinsing
operation. Release the
Prime Key
the pump returns to its primary
status.
CCW/STOP/CW Switch
In manual control mode, pull the CCW/STOP/CW switch in left
position the pump rotates counter clockwise and in right position
the pump rotates clockwise and in middle position the pump
stops.
The CCW/STOP/CW switch must be pulled out first, then put it in
the right position. Or the CCW/STOP/CW switch can not be
triggered.
Power Switch
Open or shut down the power supply.
WT600-2J OPERATING MANUAL
WT600-2J OPERATING MANUAL