Operator's Guide

5-12

Module 7005

(Calibration of the detector will set the head power, the detection mode and the compensation
value to minimise the product signal for your product - please refer to the Principles of Operation
Section if you would like to gain a better understanding of this).

To set up other products, up to one hundred in all, repeat the above process with the product number
set as required. The product number can either be changed  by using the Change product menu (see
below), or by altering the product number in the Product Setup menu and initiating a full calibration
from the Product Setup menu.

Changing products

Superscan Micro can be pre-calibrated for up to one hundred different products, which will all
initially be setup using the above new product setup procedure. Once you have set up two or more
products, you will wish to switch between the two as the product in production changes. This is
achieved using the product [CHANGE] key:

1

From any menu position press the product [CHANGE] key. The display will now show the
current product number selected and the name of the product.

2

Use the [

] and [

] keys to move up and down the list of one hundred products.

3

Once the display is showing the desired product, press the [ENTER] key to accept that choice.
The display will indicate that the product been changed and then return to the top of the menu
map, showing the product signal. Refer to the Change menu in the A - Z section for details of any
warning messages given.

Recalibrating the detector

In the course of operating the detector, you may find that certain situations require the calibration
details of the product to be altered. Examples may include where the temperature or ingredients of the
product have changed, or where an existing product number is to be used for a new product.

Two sorts of calibration cycles are available to cater for all eventualities, Full Calibrate and Update
Calibrate. Update Calibration will fine tune the compensation value (see Principles of Operation) to
minimise the product signal whilst still checking for metal in the product, whereas Full calibration
will completely calibrate the detector which necessitates turning  detection for metal off during the
calibration cycle.

Use Update Calibration to minimise the product signal when the production line is running if false
rejects are occurring because the product signal has “drifted”.

Use Full Calibration in all other circumstances, or when an Update fails to produce the required
results.

When PV Test is in use, recalibrating the detector using either Update or Full Calibration will not
reset the False Threshold value.

Update Calibration can be accessed from the Calibrate menu only, whereas Full Calibration can be
accessed from the Calibrate and the Product Setup menus.

Содержание Superscan Micro

Страница 1: ...Calibrate and Results These menus can all be accessed via the dedicated keys on the User Interface To become completely familiar with the operation of the LOMA Superscan Micro it is necessary that th...

Страница 2: ...Key Up Key Main Menu Keys Each of the four main menu keys has an LED which shows the main menu that is being accessed at any time Status LEDs The Run Detect LED indicates that the detector is searchin...

Страница 3: ...X PACK COUNT RESET BATCH STATUS REPORT RUN PV TEST T xxxxx S xxxxx LOCK KEYBOARD CHANGE SETUP CALIBRATE RESULTS SCROLL SELECTION KEYCODE 1234 CALENDAR SERVICE CHANGE 1234 PV TEST 1234 PRODUCT NO _ _ N...

Страница 4: ...note that an is shown on the display in which case go to step four If the display shows neither of these texts push the CANCEL key until it does and then go to step two or four depending upon the disp...

Страница 5: ...holds giving greater sensitivity and larger ones less sensitivity To adjust this threshold press the ENTER key The display will now show T 1000 SIG 255 Notice how the has moved to the left hand side o...

Страница 6: ...in access to the items within them The Disable Reject is in fact a command and as such is denoted with the character Within the Setup menu there are no items of data but if the Product Setup sub menu...

Страница 7: ...ing example Step One Suppose you with to alter the time of day say from 9 00am to 10 00am Use the keyboard as described above to move to the Hours parameter The display should show HOURS 9 Notice how...

Страница 8: ...f this is shown in the next section on how to set up a product for the first time where the Full Calibrate command is invoked To summarise Use the main menu keys to move into one of the four menus Use...

Страница 9: ...ou have calibrated the machine with a product which you believe has no product effect because it contains no moisture or conductive ingredients iron additives but the unit fails to achieve the desired...

Страница 10: ...Press the down key to move to the display PV TEST SETUP This menu only appears if the option is enabled via the OPTIONS menu PV TEST SETUP contain five system relative entries and four product relati...

Страница 11: ...icate that the product been changed and then return to the top of the menu map showing the product signal Refer to the Change menu in the A Z section for details of any warning messages given Recalibr...

Страница 12: ...the display indicates that the calibration cycle is complete Setting up a new product Press the SETUP key once Press the ENTER key twice Use the and keys to select the desired product number Press th...

Отзывы: