![LogTag UTREL30-WIFI Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/logtag/utrel30-wifi/utrel30-wifi_product-user-manual_1938025049.webp)
Starting the Logger
UTREL30-WiFi User Guide
49
Once
READY
disappears, release the button within
two seconds.
STARTING
will now disappear, and the
symbol will be shown. The UTREL30-WiFi now
records temperature data.
The logger will
not
start if
l
you release the button before
READY
disappears;
l
you keep holding the button for more than 2
seconds after
READY
disappears; or
l
the AAA backup batteries are critically low and
the logger is not connected to power.
The UTREL30-WiFi now automatically performs a Connection Test
wireless and network connection on page 34)
.
Push Button Start with Start Delay
If the logger has been configured for a push button
start with a start delay, the word
STARTING DELAY
is shown instead of the
symbol.
The delay time is shown in hours and minutes. The
time counts down and the UTREL30-WiFi starts
recording when it reaches
0
:
00
.
As with a regular push button start without delay, the UTREL30-WiFi will perform a
Connection Test
as soon as the recording has started. To check the connection before
the delay value has elapsed, you can initiate the connection test yourself. Please refer to
Testing your wireless and network connection (on page 34)
The timer can be cancelled and the logger reset to
READY
.
While
STARTING
and
DELAY
are shown, press and
hold the
START
/CLEAR/STOP
button.
STOPPED
will now
show. Release the button when
STARTING
disappears.
STOPPED
and
DELAY
will both disappear, and
READY
will show, together with the current time,