www.logiquip.com
|
17
INSTALLING HEPA FAN(S) & FLOW SENSOR(S)
This section applies to all models.
1. Insert the grill duct (Fig. 10) from inside the glass Cabinet and secure using four (4)
8
/
32
x ¾ screws. Ensure the flow
sensor, as applicable, is facing towards the front of the Cabinet.
2. Install the duct over the grill duct (Fig. 11). Secure the connection with the hose clamp and a flat head screwdriver.
3. Cover any unused, vacant grill duct cut-out locations with blank cover (Fig. 12).
4. Remove the HEPA Fan from its packaging, Turn the HEPA Fan switch to the ON position (marked by an ‘I’ symbol) and
then take the other end of the duct and place it over the HEPA Fans’ duct collar. Secure the duct by tightening the hose
clamp using a flat head screwdriver.
5. SIDE LOAD: Place the HEPA Fan on the bottom shelf of the component Cabinet with the ON/OFF switch facing
towards the left (Fig. 13).
SIDE LOAD ADD-ON CABINET: The second HEPA Fan can be placed on the second shelf from the bottom, and steps
repeated for the second grill duct.
TOP LOAD: Place the HEPA Fan in the top compartment shelf with the ON/OFF switch facing towards the left.
6. Connect the HEPA Fan power cord(s) into the power bar in the serviceable component area.
7. HEPA & HEPA CHANNEL PURGE ONLY: Connect the Cabinet Light Bar(s) into the power bar in the serviceable
component area.
8. Verify that the power bar is switched to the ON position.
Fig. 10
Fig. 11
Fig. 12
Fig. 13
Fig. 14
Light Bar installation for Tracking (V1) Cabinets is completed within instruction for
Electrical Plate and CPU Installation
.
Light Bar installation for Tracking (V2) Cabinets is completed within instruction for
Major Control Box Installation
.