38
ENGLISH
1. Power On
Press "START/STOP” button until a beep sounds. The LCD screen will appear for one second as unit
performs a quick diagnosis. A long tone indicates device is ready for testing.
NOTE:
Unit will not function if residual air from previous testing is present in cuff. The LCD will flash
" " until pressure is stabilized.
2. Pressurization
Initial pressure is first pumped to 190mmHg. If the current user's systolic blood pressure is over
190mmHg, the unit will automatically re-inflate to the proper shelf
NOTE:
Pressurization will gradually subside and ultimately stop when cuff is not properly applied to
the wrist. If this occurs, pressing the "START/STOP” button to turn the unit off.
3. Testing
After cuff inflation, air will slowly subside as indicated by the corresponding cuff pressure value. A
flashing "
" will appear simultaneously on screen signaling heart beat detection.
NOTE:
Remain relaxed during testing. Avoid speaking or moving body parts