LOGAN FRIABILITY TESTING SYSTEM
FAB-2S FRIABILITY TESTER
Page
9
of
12
C.
Press “Set” button, the second to last digit will start flashing.
D.
Press “Enter” button to increase the flashing digit. (Skip step C if no change is required.)
E.
To change the remaining digits repeat the above procedure.
5.5
To Reset the Friability Tester
Press “reset” button. FAB-2S will go back to default starting mode. Please note that default
mode for FAB-2S is Timer Mode. To select Counter Mode you must press “Timer/Counter”
button.
5.6
To Start Test
Once you have finished setting your time or count, press “Start/Stop” button. FAB-2S will start
rotation for timer or counter mode.
5.7
To Stop Test
FAB-2S will automatically stop when it reaches the preset time or count. To stop test before it
reaches its preset time or count, press “Start/Stop” button.
Warning:
Below information for the lab manager only, do not show technician how to change the rotation
speed to avoid future problems.
F.
To change the rotation speed:
5.8
Turn on the POWER switch while pressing and holding the “Start/Stop” button, the display
shows the last rotation speed.
5.9
Press “Set” and “Enter” button to input the new rotation speed. Proceed as to set the
Revolution Count. The speed setting is from 20-70 RPM.
5.10
Press “Reset” button into test mode for the test.
0:00:0
0