Страница 1: ... 19 C School House Rd Somerset NJ 08873 Tel 732 302 9888 Fax 732 302 9898 E Mail info loganinstruments com www loganinstruments com FAB 2S FRIABILITY TESTER with variable speed setting Instruction Manual Rev 0 Date 5 26 2015 ...

Страница 2: ...TER Page 2 of 12 TABLE OF CONTENTS 1 0 SPECIFICATIONS 2 0 DESCRIPTION OF FAB 2S FRIABILITY TESTER 3 0 DESCRIPTION OF CONTROLS 4 0 INSTALLATION AND SET UP 5 0 OPERATING PROCECURES 6 0 MAINTENANCE 7 0 TROUBLE SHOTTING 8 0 SPARE PARTS AND SERVICE INFORMATION ...

Страница 3: ... 3 of 12 1 0 SPECIFICATIONS Drum Position 2 Drum Speed Variable Speed 25 70 RPM Speed Accuracy Per Minute 1 RPM Maximum Testing Time 9 HR 59 MIN 59 SEC Maximum Count of Revolutions 99999 Revolutions Size 12 x 16 x 14 Weight 20LBS Voltage 110V 50 60Hz 220V 50 60Hz ...

Страница 4: ...les have to be dusted and weighed again Actually the difference between the weight before and after the test is determined as Friability which usually should not exceed 1 The standard Friability drum causes the samples to roll and fall during the test while the abrasion drum continuously stresses the sample by abrasion of lamellas FAB 2S Friability Tester can be used 1 or 2 drum A noiseless gear m...

Страница 5: ...LOGAN FRIABILITY TESTING SYSTEM FAB 2S FRIABILITY TESTER Page 5 of 12 2 1 FAB 2S AUTOMATED DISSOLUTION TESTER FRONT VIEW ...

Страница 6: ...ounter Timer Indicator Light Purpose To indicate Friability Tester is in timer mode Counter Indicator Light Purpose To indicate Friability Tester is in counter mode Start Stop Button Purpose To start or stop the timing or counting Counter Timer Button Purpose To select counter mode or timer mode Reset Button Purpose To reset Friability Tester to re start Enter Button Purpose To enter the digits Se...

Страница 7: ...rocedure 5 1 Turn power switch on 5 2 Select timer mode or counter mode by pressing Timer Counter button 5 3 Timer Mode With Timer indicator light on display will show a default time of 0 04 00 This means when Start Stop button has been pressed drum will rotate for 4 minutes and then stop To Set Test Time Make sure timer mode indicator light is on before proceeding Set Seconds A Press set button t...

Страница 8: ...will start flashing B Press Enter button to increase the flashing digit Skip step B if no change is required C Press Set button to end the setting process 5 4 Counter Mode With counter indicator light on display shows a default revolution count of 0 01 00 This means when Start Stop button has been pressed drum will go 100 rotations and then stop To Set Timer Mode Make sure the counter indicator li...

Страница 9: ... your time or count press Start Stop button FAB 2S will start rotation for timer or counter mode 5 7 To Stop Test FAB 2S will automatically stop when it reaches the preset time or count To stop test before it reaches its preset time or count press Start Stop button Warning Below information for the lab manager only do not show technician how to change the rotation speed to avoid future problems F ...

Страница 10: ...ast 2 digits can be programmed 25 RMP 70 RMP B Press the Set button the last digit will start flashing C Press Enter to increase the flashing digit from 0 to 9 D Press Set button and the fourth digit will start flashing E Press Enter to increase the flashing digit from 2 to 7 F Press Reset to enter the new speed setting into the FAB 2S G Turn off the FAB 2S and power on again to check the new spee...

Страница 11: ... Unit Keep on a flat and clean surface Clean the surface by using a dry cloth Do not use any solvent 6 3 Drive Shaft Make sure both shafts are dry at all times Clean shaft with dry cloth Do not use any solvent Fuse Replacement Instruction To replace the fuse turn off the power switch Gently press the fuse holder knob in and turn counter clockwise Take the old fuse out and replace it with a fuse Pa...

Страница 12: ...ck Line Cord Check Fuse When power is on constantly long beep Check Buttons if join with case No rotation Check Motor and Belt If further assistance is needed please contact your local distributor or Logan Instruments Address 19 C School House Rd Somerset NJ 08873 Tel 732 302 9888 Fax 732 302 9898 Email info loganinstruments com ...