2. Press the enter key #4 until the display shows terminal 0

3. By using the "+" and "-" keys for the grade adjustment #3 you can change the "0" behind terminal to

"1" to activate the Terminal mode.

4. Press enter #4 and "prot id" will be shown on the display, for Lode protocol this should be set to "1".

Set the protocol as described in the "Ergometer setup" chapter. 

The following protocols can be selected

Lode = Lode protocol

trAck = Trackmaster communication protocol

UUood = Woodway communication protocol

HPcoS = HP Cosmos communication protocol

gE = GE protocol

5. Press enter #4 and "baudrate" will be displayed to set the correct baudrate, select "0" for auto select.

The following baudrates can be set:

Baudrate = 0 (Autoselect dependent on protocol-id)

2400;

4800;

9600;

19200;

38400;

57600

6. Press enter #4 and "interface" will be displayed, you can choose the communication port which you

would like to select.

Interface = CIM interface choice. Select "RS232" or "USB" depending on the cable that you connected to

the treadmill.

0 = RS232

1 = RS232-AUX

2 = USB

7. Press enter #4; the next step "Neg Elev" is read only and is "1" when a negative elevation option has

been added to the treadmill, without a negative elevation option this value shows "0".

Neg Elev = Negetive elevation (read-only)

0 = not present

1 = present

Press the stop button #2 on the control unit or wait 10 seconds to leave the settings menu and let your

changes take effect.

18/22

INTERFACE SHEET - COSMED OMNIA - CORIVAL CPET / VERSION: 1.0 - 2016-11-29

Содержание CORIVAL CPET

Страница 1: ...mnia Corival CPET Service documentation Lode BV Zernikepark 16 NL 9747 AN Groningen The Netherlands t 31 50 571 28 11 f 31 50 571 67 46 ask lode nl www lode nl INTERFACE SHEET COSMED OMNIA CORIVAL CPE...

Страница 2: ...Ergometer Setup 4 Setting up the hardware 4 1 Connection overview 4 2 Connecting with the usb cable 4 3 Connecting with the RS232 cable 5 Treadmill or ergometer with 3 5 touch screen unit 6 Treadmill...

Страница 3: ...nt supported products are Quark CPET K5 Quark RMR Quark PFT This interface form covers the following Lode products All Corival ergometers starting with serialnumber 9609xx 20150001 If your treadmill o...

Страница 4: ...he Cable to the Cosmed Quark on the AUX RS232 port no 17 in picture if you are using an RS232 cable or connect the USB A to B cable directly to the computer when using the USB A to B cable In case you...

Страница 5: ...port selection Press OK to confirm and start a test Make sure the ergometer is set up correctly before pressing OK The procedure is explained in the next chapters 3 3 Ergometer Setup The following pro...

Страница 6: ...utomatically You can find the driver on the Lode website under the Services tab Locations For the Treadmill the communication ports are Located on the front side near the power switch For the bicycle...

Страница 7: ...2 Connect the cable to the Lode ergometer 3 Connect the cable to the external device 4 3 Connecting with the RS232 cable Connecting with the RS232 cable 1 Depending on the external device you may need...

Страница 8: ...like the kind of device the serial number and the connected modules is displayed After initialisation the terminal mode is automatically displayed Swipe with your finger over the screen as indicated w...

Страница 9: ...nsaver The menu Terminal Settings lets you set the communication settings for external communication The menu Service is only intended for service purposes to be executed by factory trainend personel...

Страница 10: ...Select the Communication protocol as desribed in Ergometer setup chapter Select Default for the baudrate 10 22 INTERFACE SHEET COSMED OMNIA CORIVAL CPET VERSION 1 0 2016 11 29...

Страница 11: ...232 or USB depending in which cable you have connected to the Lode ergometer Select the button with the house to go back to the main menu 11 22 INTERFACE SHEET COSMED OMNIA CORIVAL CPET VERSION 1 0 20...

Страница 12: ...Select Terminal to set the ergometer in terminal mode Your device is now ready for external control 12 22 INTERFACE SHEET COSMED OMNIA CORIVAL CPET VERSION 1 0 2016 11 29...

Страница 13: ...evice is shown on the display during boot mode After boot up the device shows the Test window or main menu To access the menu from the Test window you need to swipe the graph area of the touch screen...

Страница 14: ...n using the red marked scroll button and select Terminal Select the communication protocol ad described in the Ergometer setup chapter Use the scroll button to view the complete list 14 22 INTERFACE S...

Страница 15: ...Select Default for the baudrate Select RS232 or USB depending on the type of cable that you connected to the Lode ergometer 15 22 INTERFACE SHEET COSMED OMNIA CORIVAL CPET VERSION 1 0 2016 11 29...

Страница 16: ...on optional Your ergometer can be set to boot up in terminal mode Follow these steps to activate this function Go to the settings menu Select Default Start Menu Select Terminal Switch the ergometer of...

Страница 17: ...7 Nummerical Control Unit Treadmill only 1 Press and hold key 1 and 2 as shown on Figure 1 for 10 seconds 17 22 INTERFACE SHEET COSMED OMNIA CORIVAL CPET VERSION 1 0 2016 11 29...

Страница 18: ...r auto select The following baudrates can be set Baudrate 0 Autoselect dependent on protocol id 2400 4800 9600 19200 38400 57600 6 Press enter 4 and interface will be displayed you can choose the comm...

Страница 19: ...can be downloaded from the Lode website if you are a registered user http lode nl en product getDownload 469 Connect an USB cable from the computer to the USB port of the treadmill or use a RS232 cabl...

Страница 20: ...nected 3 Select preset Lode Bus Service Protocol 4 Press OK 5 Press Break 6 Switch the device off wait 10 seconds 7 Switch the device on The screen as shown left should show up within 5 seconds after...

Страница 21: ...as described in the Ergometer set up chapter Choose which port you want to use RS232 0 or USB 1 This setting refers to the type of cable that you have connected to the Lode ergometer Select default fo...

Страница 22: ...ct Disable 0 for Global communication timeout Press Y to save the new settings and restart the device Your device is now ready for use with the external controller Note that the device is automaticall...

Отзывы: