9 Maintenances
All of the maintenance operations that involve handling parts of the unit must only be carried out by
suitably trained personnel and in full compliance with all safety standards.
Moreover, these operations must only be carried out with the unit off, unplugged and electrically
isolated.
All the described intervals are intended with a use and storage in an environment with features like
ones described in 2.3
Always follow the maintenances intervals described in this manual.
Not respect them may result in an incorrect operation of the unit and any damage will not covered by
warranty.
This section describes some checks, cleaning and maintenance operations required to ensure correct
operation of the appliance.
In these cases, the system activates pre-alarm messages (nitrogen production still continues) to remind the
user to complete the maintenance operation.
The described intervals refers to the unit usage with general specification related to the ones described in the
table (2.3.6) and operation at 80% of its maximum power.
For any other maintenance and/or service operations, contact the reseller and/or manufacturer.
9.1.1
Preventive checks and maintenances
Operation
Description
Interval
Check
Check the condensate drain system
24 hours
Check
Check for abnormal noise and vibrations
24 hours
Check/
Maintenance
Check the air intake filter and clean if
necessary
After the first 400 hours / 2 months.
Subsequently every 2500 hours / 1 year